INSET
Report
- Report Number
- 3003442380-2025-15688
- Event Type
- Malfunction
- Date Received
- November 3, 2025
- Date of Event
- October 6, 2025
- Report Date
- October 28, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (3003442380-2025-15688), WAS SUBMITTED ON 03-NOV-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 18-MAR-2025. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6012424, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012424 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 121 IN THE LINE 10, ON 18/MAR/2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5C02463 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 02 AND MANUFACTURED IN THE MACHINE ITL03, ON 15-MAR-202, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B05647 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 02 AND MANUFACTURED IN THE MACHINE ITL03, ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS - H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6012424, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012424 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 121 IN THE LINE 10, ON 18/MAR/2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5C02463 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 02 AND MANUFACTURED IN THE MACHINE ITL03, ON 15-MAR-202, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B05647 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 02 AND MANUFACTURED IN THE MACHINE ITL03, ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. NOTE: PROTECTOR UP EXTENDED THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS INT THE TUBING. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2686928 | INSET | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001680 | 6012424 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male |