FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 23430024 · Received October 30, 2025

Report

Report Number
9617229-2025-18680
Event Type
Injury
Date Received
October 30, 2025
Date of Event
September 23, 2025
Report Date
November 19, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON DECEMBER 16, 2025, WITH LOT NUMBER 1242264. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE THROUGH VISUAL INSPECTION AS IT IS A PHYSIOLOGICAL PHENOMENON. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES AND DEFORMATION) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS. ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D.9., H.2., H.3., H.6.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE III.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D.6B, G2, H6.

Description of Event or Problem · 0

PATIENT REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III" AND "PAIN". LATER HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

PATIENT REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III" AND "PAIN". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

PATIENT REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III" AND "PAIN". LATER HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1617458 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1242264

Patients

Seq Age Sex Outcome Treatment
1 33 YR Female Required Intervention