FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND HIGH PROFILE

MDR report key: 23426337 · Received October 30, 2025

Report

Report Number
1645337-2025-12131
Event Type
Injury
Date Received
October 30, 2025
Date of Event
October 7, 2025
Manufacturer
MENTOR TEXAS
Product Code
FWM
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RIGHT BREAST PROSTHESIS DEFLATION, BILATERAL BREAST PTOSIS. D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 21-NOV-2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE SUSPECT MEDICAL DEVICE IS A MENTOR SMOOTH ROUND HIGH PROFILE 380CC SALINE BREAST PROSTHESIS, CATALOG #3503380, LOT #271546, PMA #P990075. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. D4: UDI: (01)GTIN UNAVAILABLE, PRODUCT MADE PRIOR TO GTIN COMPLIANCE DATE. ON 03-DEC-2025, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED DEFLATION IN THE BREAST IMPLANT AND BREAST PTOSIS. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED VISUAL INSPECTION, LEAK TESTING, AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT NO DAMAGE OR ANOMALIES WERE OBSERVED ON THE BREAST IMPLANT. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND IT IDENTIFIED A TEAR ON THE ANTERIOR VIEW, MEASURING APPROXIMATELY 0.1 CM. IN ADDITION, NO OTHER LEAK SITES WERE DETECTED DURING THE ANALYSIS. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN THE WHOLE AREA OF THE TEAR. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. BASED ON THE INFORMATION CURRENTLY AVAILABLE, MICROSCOPIC EXAMINATION OF THE RETURNED PRODUCT INDICATES THAT THE IMPLANT COULD HAVE BEEN DAMAGED DURING OR SUBSEQUENT TO IMPLANTATION. THE PRODUCT INSERT DATA SHEET CAUTIONS TO NOT ALLOW CAUTERY DEVICES OR SHARP INSTRUMENTS, SUCH AS SCALPELS, SUTURE NEEDLES, HYPODERMIC NEEDLES, HEMOSTATS, ADSON FORCEPS OR SCISSORS TO CONTACT THE DEVICE DURING THE IMPLANTATION OR OTHER SURGICAL PROCEDURES. SOME BREAST PTOSIS IS A NORMAL COMPONENT OF THE MATURE BREAST. PTOSIS BECOMES UNDESIRABLE WHEN THE BREAST PARENCHYMA PREDOMINATES BELOW THE AREOLA, DROOPS CONSIDERABLY BELOW THE INFRAMAMMARY FOLD AND THE NIPPLE POINTS DOWNWARD. AUGMENTATION ALONE, WITHOUT CONSIDERATION OF THE PTOSIS CAN PRODUCE A LESS THAN DESIRABLE COSMETIC RESULT KNOWN AS A "ROCK IN SOCK" DEFORMITY OR AN INCREASED PTOTIC APPEARANCE. PTOSIS IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER:(B)(4).

Additional Manufacturer Narrative · 0

ON 31-OCT-2025, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. THE INVESTIGATION WAS COMPLETED ON A PHOTO OF THE SUSPECT MEDICAL DEVICE BECAUSE THE PHYSICAL DEVICE HAS NOT BEEN RECEIVED BY MENTOR. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A DEFLATION IN THE SALINE BREAST IMPLANT AND DEVELOPED BREAST PTOSIS. UPON VISUAL EVALUATION OF THE IMAGE PROVIDED IN THE COMPLAINT, THE IMPLANT WAS OBSERVED TO BE DEFLATED. AS THE PRODUCT INVOLVED IN THIS COMPLAINT WAS NOT RECEIVED, THE DEVICE COULD NOT BE ANALYZED ACCORDING TO OUR PROCEDURES. SOME BREAST PTOSIS IS A NORMAL COMPONENT OF THE MATURE BREAST. PTOSIS BECOMES UNDESIRABLE WHEN THE BREAST PARENCHYMA PREDOMINATES BELOW THE AREOLA, DROOPS CONSIDERABLY BELOW THE INFRAMAMMARY FOLD AND THE NIPPLE POINTS DOWNWARD. AUGMENTATION ALONE, WITHOUT CONSIDERATION OF THE PTOSIS CAN PRODUCE A LESS THAN DESIRABLE COSMETIC RESULT KNOWN AS A "ROCK IN SOCK" DEFORMITY OR AN INCREASED PTOTIC APPEARANCE. PTOSIS IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH UNSPECIFIED MENTOR SMOOTH SALINE BREAST PROSTHESES WHEN THE RIGHT BREAST PROSTHESIS DEFLATED POST IMPLANTATION. ADDITIONALLY, THE PATIENT DEVELOPED BILATERAL BREAST PTOSIS. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL BOOST BREAST IMPLANT 325CC GEL BREAST PROSTHESES ON (B)(6) 2025. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT BREAST PROSTHESIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2380931 MENTOR SMOOTH ROUND HIGH PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 271546

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention