FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2342611 · Received November 18, 2011

Report

Report Number
2050012-2011-07628
Event Type
Malfunction
Date Received
November 18, 2011
Date of Event
October 27, 2011
Report Date
October 27, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ROOT CAUSE IS UNKNOWN; HOWEVER, THE ISSUE WAS RESOLVED AFTER THE FIELD SERVICE ENGINEER PERFORMED THE SERVICES. ADDITIONAL MDRS HAVE BEEN FILED BASED ON THE SAME SET OF EVENTS: MEDWATCH #2050012-2011-07626 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). MEDWATCH #2050012-2011-07627 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). (NOTE: THE BECKMAN COULTER, INC. IDENTIFIER FOR THIS REPORT IS (B)(4).)

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT THAT THE UNICEL DXC 600 SYNCHRON SYSTEM GENERATED ERRATIC QUALITY CONTROL RESULTS FOR TRIGLYCERIDE (TG) AND MAGNESIUM (MG) ON TWO INSTRUMENTS. THIS REPORT IS BEING FILED ON ONE OF THE INSTRUMENTS, SERIAL NUMBER (B)(4). PLEASE NOTE THAT ANOTHER REPORT WAS FILED FOR THE OTHER INSTRUMENT, SERIAL NUMBER (B)(4), AS MEDWATCH #2050012-2011-07627 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). CUSTOMER STATED THAT TG VALUES WERE >500 MILLIGRAMS PER DECILITER (MG/DL), AND UPON REPEAT WAS <120 MG/DL. ALSO, THE MG VALUES WERE >5.0 MG/DL, AND UPON REPEAT WAS 2.0 MG/DL. THE CUSTOMER TECHNICAL SPECIALIST (CTS) SUGGESTED TO CUSTOMER THAT THE ROOT CAUSE MIGHT BE THE LABORATORY'S DEIONIZED WATER SYSTEM, AS THE TWO INSTRUMENTS SHARE THE SAME WATER SYSTEM. CUSTOMER EXPLAINED THAT SERVICE ON THE WATER SYSTEM WAS DUE, AND THAT CUSTOMER WOULD HAVE THE TANKS CHANGED, ALONG WITH THE .2 MICRON FILTERS. CUSTOMER SUSPECTS THAT THESE HAVE NOT BEEN CHANGED FOR OVER A YEAR. CUSTOMER REQUESTED SERVICE TO ENSURE THAT THE ISSUES WOULD BE RESOLVED. CUSTOMER STATED THAT PATIENT TREATMENT WAS NOT AFFECTED. ON THE SAME DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND DETERMINED THAT THE INSTRUMENT WAS GENERATING SUPPRESSED TRIGLYCERIDE RESULTS. AFTER REPLACING BOTH THE REAGENT PROBES AND THE WASH COLLARS, THE FSE FOUND THAT THE CARTRIDGE CHEMISTRY MIXER WASH STATION WAS NOT PROPERLY FUNCTIONING. THE FSE THEN REPLACED ITS VALVE, AFTER WHICH INSTRUMENT PERFORMANCE WAS VERIFIED TO ENSURE THAT THE SERVICES PERFORMED EFFECTIVELY RESOLVED THE ISSUE. THE FSE THEN VERIFIED THAT THE PRECISION RUN AND QUALITY CONTROLS WERE ACCEPTABLE AND WITHIN RANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER, INC. DXC 600 PRO

Patients

Seq Age Sex Outcome Treatment
1