UNICEL DXC 600 SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2011-07628
- Event Type
- Malfunction
- Date Received
- November 18, 2011
- Date of Event
- October 27, 2011
- Report Date
- October 27, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE ROOT CAUSE IS UNKNOWN; HOWEVER, THE ISSUE WAS RESOLVED AFTER THE FIELD SERVICE ENGINEER PERFORMED THE SERVICES. ADDITIONAL MDRS HAVE BEEN FILED BASED ON THE SAME SET OF EVENTS: MEDWATCH #2050012-2011-07626 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). MEDWATCH #2050012-2011-07627 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). (NOTE: THE BECKMAN COULTER, INC. IDENTIFIER FOR THIS REPORT IS (B)(4).)
CUSTOMER CALLED TO REPORT THAT THE UNICEL DXC 600 SYNCHRON SYSTEM GENERATED ERRATIC QUALITY CONTROL RESULTS FOR TRIGLYCERIDE (TG) AND MAGNESIUM (MG) ON TWO INSTRUMENTS. THIS REPORT IS BEING FILED ON ONE OF THE INSTRUMENTS, SERIAL NUMBER (B)(4). PLEASE NOTE THAT ANOTHER REPORT WAS FILED FOR THE OTHER INSTRUMENT, SERIAL NUMBER (B)(4), AS MEDWATCH #2050012-2011-07627 (BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4)). CUSTOMER STATED THAT TG VALUES WERE >500 MILLIGRAMS PER DECILITER (MG/DL), AND UPON REPEAT WAS <120 MG/DL. ALSO, THE MG VALUES WERE >5.0 MG/DL, AND UPON REPEAT WAS 2.0 MG/DL. THE CUSTOMER TECHNICAL SPECIALIST (CTS) SUGGESTED TO CUSTOMER THAT THE ROOT CAUSE MIGHT BE THE LABORATORY'S DEIONIZED WATER SYSTEM, AS THE TWO INSTRUMENTS SHARE THE SAME WATER SYSTEM. CUSTOMER EXPLAINED THAT SERVICE ON THE WATER SYSTEM WAS DUE, AND THAT CUSTOMER WOULD HAVE THE TANKS CHANGED, ALONG WITH THE .2 MICRON FILTERS. CUSTOMER SUSPECTS THAT THESE HAVE NOT BEEN CHANGED FOR OVER A YEAR. CUSTOMER REQUESTED SERVICE TO ENSURE THAT THE ISSUES WOULD BE RESOLVED. CUSTOMER STATED THAT PATIENT TREATMENT WAS NOT AFFECTED. ON THE SAME DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND DETERMINED THAT THE INSTRUMENT WAS GENERATING SUPPRESSED TRIGLYCERIDE RESULTS. AFTER REPLACING BOTH THE REAGENT PROBES AND THE WASH COLLARS, THE FSE FOUND THAT THE CARTRIDGE CHEMISTRY MIXER WASH STATION WAS NOT PROPERLY FUNCTIONING. THE FSE THEN REPLACED ITS VALVE, AFTER WHICH INSTRUMENT PERFORMANCE WAS VERIFIED TO ENSURE THAT THE SERVICES PERFORMED EFFECTIVELY RESOLVED THE ISSUE. THE FSE THEN VERIFIED THAT THE PRECISION RUN AND QUALITY CONTROLS WERE ACCEPTABLE AND WITHIN RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 600 PRO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |