FDA Adverse Event Injury Summary report: N

QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8

MDR report key: 23422771 · Received October 30, 2025

Report

Report Number
3016746283-2025-00014
Event Type
Injury
Date Received
October 30, 2025
Date of Event
October 1, 2025
Report Date
May 11, 2026
Manufacturer
SPECTRUM MEDICAL S.R.L.
Product Code
DTR
UDI-DI
08051160301171
PMA / PMN Number
K223879
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RELATED TO THE EVENT WILL BE AVAILABLE AFTER COMPLETION OF INVESTIGATION ACTIVITIES, WHICH WOULD BE ADDRESSED IN A FOLLOW-UP REPORT.

Additional Manufacturer Narrative · 0

PRELIMINARY ANALYSIS PERFORMED REVEALED A COMBINATION OF TWO HOLES OR ONE HOLE AND ONE DEFECT, IN A SYMMETRICAL POSITION AND IN AN AREA WHERE THE KNURLED PATTERN ANGLE CHANGES. AFTER PRELIMINARY ANALYSIS, THE ROOT CAUSE OF THE PROBLEM HAS NOT YET BEEN IDENTIFIED. ADDITIONAL TESTS ARE ONGOING. CONSIDERING THE POTENTIAL IMPACT ON THE SAFETY OF THE PATIENT, FLUIDIC TESTING CONFIRMED THAT IN CASE OF UNWANTED LOSS OF METAL SHEET INTEGRITY AND UNDER WORST-CASE CONDITIONS, FLOW OCCURS FROM BLOOD TO HTF, ENSURING THAT NO HTF CAN ENTER THE BLOOD COMPARTMENT. ADDITIONAL INFORMATION WILL BE DISCUSSED IN A FOLLOW UP REPORT.

Additional Manufacturer Narrative · 0

PRELIMINARY FINDINGS AVAILABLE IN CURRENT FOLLOW-UP 001 REPORT: VISUAL INSPECTION AND DEFECT IDENTIFICATION: TWO SYMMETRICAL HOLES WERE FOUND IN THE CENTRAL AREA OF THE DEVICE WHERE THE KNURLED PATTERN CHANGES DIRECTION. SCANNING ELECTRON MICROSCOPY (SEM): SEM IMAGING WAS USED TO CHARACTERIZE THE DIMENSIONS OF THE HOLES AND TO OBTAIN DETAILS OF THE METAL SHEET SURFACE NEAR THE HOLES. FLUID TEST RESULTS: TESTING CONFIRMED THAT EVEN WITH THE HOLES PRESENT, UNDER WORST-CASE CONDITIONS, FLOW OCCURS FROM BLOOD TO HTF, ENSURING THAT NO HTF CAN ENTER THE BLOOD COMPARTMENT. THIS IS DUE TO THE FROSTY DESIGN: HTF IS DRAWN OUT OF THE HEAT EXCHANGER, SO THE PRESSURE INSIDE THE HTF COMPARTMENT IS ALWAYS NEGATIVE, WHILE THE BLOOD COMPARTMENT IS ALWAYS UNDER POSITIVE PRESSURE. AS A RESULT, THE PRESSURE DIFFERENTIAL ACROSS THE METAL SHEET GUARANTEES THAT ONLY BLOOD CAN ENTER THE HTF COMPARTMENT, NOT VICE VERSA. FURTHER TESTING IS ONGOING TO GATHER MORE DETAILS THAT WILL HELP IDENTIFY THE ROOT CAUSE OF THE PROBLEM. THESE TESTS INCLUDE: ANALYSIS OF METAL SHEETS PERFORMED BY AN EXTERNAL LAB. STUDY OF CURRENT/VOLTAGE VERSUS HOLE FORMATION. A FINAL AND COMPLETE REPORT WILL BE RELEASED WHEN ALL THE RESULTS WILL BE AVAILABLE.

Description of Event or Problem · 0

BLOOD TO GLYCOL LEAK ON HX. CCPS DEPRIMED HX AND COULD SEE VISIBLE BLOOD LEAK. HX WENT "DARK" AFTER MINUTES ON CPB. NO CHANGES WERE NOTED TO PRIMING PROTOCOL PER CHIEF PERFUSIONIST.

Description of Event or Problem · 0

BLOOD TO GLYCOL LEAK ON HX. CCPS DEPRIMED HX AND COULD SEE VISIBLE BLOOD LEAK. HX WENT "DARK" AFTER MINUTES ON CPB. NO CHANGES WERE NOTED TO PRIMING PROTOCOL PER CHIEF PERFUSIONIST.

Description of Event or Problem · 0

BLOOD TO GLYCOL LEAK ON HX. CCPS DEPRIMED HX AND COULD SEE VISIBLE BLOOD LEAK. HX WENT "DARK" AFTER MINUTES ON CPB. NO CHANGES WERE NOTED TO PRIMING PROTOCOL PER CHIEF PERFUSIONIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321704 QUANTUM PUREFLOW STANDARD HEAT EXCHANGER HIGH FLOW 3/8 STANDARD HEAT EXCHANGER DTR SPECTRUM MEDICAL S.R.L. HX55W-S0 H000059 08051160301171

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention