QUICK SET
Report
- Report Number
- 3003442380-2025-15504
- Event Type
- Injury
- Date Received
- October 30, 2025
- Date of Event
- September 26, 2025
- Report Date
- October 22, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244017450
- PMA / PMN Number
- K160648
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011034, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 17-OCT-2025 AGAINST "FINAL REPORTING DECISION" EQUAL "SERIOUS INJURY AND DEATH", "LOT NUMBER" CRITERIA EQUAL LOT NUMBER "6011034". THE COUNT OF COMPLAINT IS 6 WHICH EXCEEDS 3. FURTHER INVESTIGATION IS REQUIRED VIA CORRECTIVE AND PREVENTIVE ACTION (CAPA) DETERMINATION ASSESSMENT USING STATISTICAL ANALYSIS. THE COMPLAINTS NUMBER ARE: (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011034 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURED IN THE MULTIVAC 12 ON 11-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5A01035 MANUFACTURED IN THE MACHINE 04, 08, ON 06-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. GLUING OF TUBING OF THE LOT 5A01036 MANUFACTURED IN THE MACHINE 04, 08, ON 07-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. GLUING OF TUBING OF THE LOT 4M00061 MANUFACTURED IN THE MACHINE 04, 08, ON 08-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. GLUING OF TUBING OF THE LOT 4M02678 MANUFACTURED IN THE MACHINE 04, 08, ON 18-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED, THEREFORE, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. VISUAL TEST ACCORDING TO WORK INSTRUCTION (WI) VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR FLOW ACCORDING TO (WI) VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST, FOR THE CODE OCCLUSION (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED). NO T/S, INCONCLUSIVE T/S, OR PARTIAL T/S DONE, REFER TO CANNOT BE DETERMINED CANNOT BE DETERMINED. NO ESCALATION REQUIRED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO DEFECT ON TEST OF REFERENCE SAMPLES. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, THE THRESHOLDS OF 6 REPORTABLE COMPLAINTS ARE MET FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT REQUIRES FURTHER CORRECTIVE AND PREVENTIVE ACTION (CAPA) DETERMINATION ASSESSMENT.
REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON 26-SEP-2025 AND EVENTUALLY HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS AND POSITIVE KETONES. IT WAS STATED THAT INSULIN ABSORPTION CAN BE A POSSIBILITY INSULIN FLOW BLOCK. THE PATIENT FELT SICK, CONFUSION AND EXPERIENCED EXTREME PAIN. THE BLOOD GLUCOSE LEVEL WAS 800 MG/DL AT THE TIME OF HOSPITALIZATION AND THE PATIENT GOT TREATED WITH INTRAVENOUS. THE PATIENT WAS IN THE HOSPITAL FOR MORE THAN 24 HOURS. THE PATIENT TREATED HIGH BLOOD GLUCOSE WITH MANUAL INJECTION AND INSULIN PEN. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1395395 | QUICK SET | UNO QUICK-SET 60/6 SC1 MECA | FPA | UNOMEDICAL A/S | MMT-399A | 6011034 | 05705244017450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Hospitalization| R |