FDA Adverse Event Malfunction Summary report: N

TANDEM DIABETES 3ML LL WITH CO-PACKED NEEDLE 26GX3/8"KDL

MDR report key: 23418518 · Received October 29, 2025

Report

Report Number
1035907-2025-00037
Event Type
Malfunction
Date Received
October 29, 2025
Date of Event
July 19, 2025
Report Date
February 11, 2026
Manufacturer
EXELINT INTERNATIONA, CO
Product Code
FMF
UDI-DI
00020221261158
PMA / PMN Number
K861153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION OF THE ISSUE THAT INCLUDED THE FOLLOWING: RETAINED SAMPLES WERE EXAMINED THAT INCLUDED THE FOLLOWING: FIVE RANDOMLY RETAINED SAMPLES WERE SELECTED FROM THIS SPECIFIC BATCH FOR SIMULATED ASPIRATION AND INJECTION TESTS. NO ISSUES WITH ASPIRATION AND INJECTION WERE FOUND. ADDITIONALLY, A REVIEW OF THE PRODUCTION PROCESS, FINISHED PRODUCT INSPECTION, AND INCOMING INSPECTION F THE INJECTION NEEDLES WERE CONDUCTED, AND NO ABNORMALITIES WERE IDENTIFIED IN THE INJECTION NEEDLES. PRODUCT RECORDS - THE RECORDS RELATED TO PRODUCT 1001215 LOT 240508, PRODUCT RECORDS AND INSPECTION REPORT WERE REVIEWED. ALL DOCUMENTS WERE FOUND TO BE ACCEPTABLE, AND NO NONCONFORMITIES. BASED ON THE INVESTIGATION, THE REPORTED ISSUE COULD NOT BE REPRODUCED. NO DEFECTS OR ABNORMALITIES WERE FOUND IN THE RETAINED SAMPLES, AND NO NONCONFORMITIES WERE IDENTIFIED. (B)(4).

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

CALLER REPORTED THEY COULD NOT DRAW AIR IN OR OUT OF THE SYRINGE. WHILE TRYING TO DO SO CUSTOMER ACCIDENTALLY STABBED THEIR FINGER WITH THE NEED AND INJECTED AIR INTO THEMSELVES. CUSTOMER WENT TO THE ER AND DOCTOR DID NOT HAVE TO TAKE ANY ACTION UPON OBSERVATION. NO SUBSTANTIAL DAMAGE OCCURRED TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2685819 TANDEM DIABETES 3ML LL WITH CO-PACKED NEEDLE 26GX3/8"KDL SYRINGE WITH NEEDLE FMF EXELINT INTERNATIONA, CO 1001215 240508 00020221261158

Patients

Seq Age Sex Outcome Treatment
1 NA Female Hospitalization