FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 13MM

MDR report key: 23418093 · Received October 29, 2025

Report

Report Number
0001038806-2025-02808
Event Type
Injury
Date Received
October 29, 2025
Date of Event
September 18, 2025
Report Date
January 5, 2026
Manufacturer
ZIMVIE US CORP LLC
Product Code
DZE
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). MULTIPLE REPORTS ARE BEING SUBMITTED FOR THIS EVENT. A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. D4: DEVICE UDI NUMBER UNKNOWN / NOT PROVIDED. G4: PMA/510(K) NUMBERS: K011028, K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D3: EMAIL ADDRESS. D8-9: DEVICE SERVICE AND AVAILABILITY. G1: CONTACT OFFICE - FIRST/GIVEN NAME, LAST NAME AND EMAIL ADDRESS. G1: CONTACT OFFICE - MANUFACTURER SITE - EMAIL ADDRESS. G3: DATE RECEIVED BY MANUFACTURER . G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 0201146. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 0201146 FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿PERI-IMPLANTITIS¿. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED LOSS OF INTEGRATION DUE TO PERI-IMPLANTITIS. TOOTH SITES #22, #23, #24, AND #25. IMPLANTS WERE REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2741871 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 13MM DENTAL IMPLANT DZE ZIMVIE US CORP LLC 0201146

Patients

Seq Age Sex Outcome Treatment
1 92 YR Unknown Required Intervention