IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 13MM
Report
- Report Number
- 0001038806-2025-02806
- Event Type
- Injury
- Date Received
- October 29, 2025
- Date of Event
- September 18, 2025
- Report Date
- January 5, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). MULTIPLE REPORTS ARE BEING SUBMITTED FOR THIS EVENT. A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. D4: DEVICE UDI NUMBER UNKNOWN / NOT PROVIDED. G4: PMA/510(K) NUMBERS: K011028, K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
ZIMVIE COMPLAINT NUMBER (B)(4). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 0201146. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 0201146 FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿PERI-IMPLANTITIS¿ A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED LOSS OF INTEGRATION DUE TO PERI-IMPLANTITIS. TOOTH SITES #22, #23, #24, AND #25. IMPLANTS WERE REMOVED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 171068 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 13MM | DENTAL IMPLANT | DZE | ZIMVIE US CORP LLC | 0201146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 92 YR | Unknown | Required Intervention |