FDA Adverse Event Injury Summary report: N

ELITE IQ

MDR report key: 23415300 · Received October 29, 2025

Report

Report Number
1222993-2025-00051
Event Type
Injury
Date Received
October 29, 2025
Date of Event
October 3, 2025
Report Date
October 29, 2025
Manufacturer
EL. EN. ELECTRONIC ENGINEERING SPA
Product Code
GEX
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT EXPERIENCED BURNS ON THE LEG FOLLOWING A LASER HAIR REDUCTION TREATMENT USING ELITE IQ. CYNOSURE CLINICAL REACHED OUT TO THE SITE FOR ADDITIONAL INFORMATION. IT WAS DETERMINED THE USER ERROR WAS INVOLVED. OPERATOR ADMITS THAT PATIENTLY RECENTLY WENT ON A TRIP TO GREECE AND LEGS WERE STILL TAN FROM THE TRIP. OPERATOR RELAYED THEY STARTED WITH THE YAG, BUT THEN DECIDED TO CHANGE AND USE ALEX WAVELENGTH INSTEAD. PER CYNOSURE CLINICAL, "THE ALEX WAVELENGTH IS NOT APPROPRIATE FOR TANNED SKIN." IMAGES WERE PROVIDED BY THE SITE AND GIVEN TO CYNOSURE MEDICAL DIRECTOR FOR CONSULTATION. CYNOSURE MEDICAL DIRECTOR STATES THAT THE AREA HAS A FULL THICKNESS BURN AND WILL SCAR. PATIENT WAS ALSO GIVEN BIAFINE CREAM AS PREVENTATIVE MEDICATION. A SERVICE EVALUATION WAS PERFORMED BY CYNOSURE FIELD SERVICE ENGINEER (FSE) WHO CONFIRMED THE DEVICE'S OUTPUTS WERE WITHIN SPECIFICATION. BURNS ARE AN EXPECTED SIDE EFFECT FROM SUCH TREATMENT. SINCE A FULL THICKNESS BURN OCCURRED FROM THIS EVENT, THIS IS CONSIDERED A SERIOUS INJURY AND THEREFORE REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212440 ELITE IQ ELITE IQ GEX EL. EN. ELECTRONIC ENGINEERING SPA

Patients

Seq Age Sex Outcome Treatment
1 NA Female Other