QUICK SET
Report
- Report Number
- 3003442380-2025-15424
- Event Type
- Malfunction
- Date Received
- October 29, 2025
- Date of Event
- October 1, 2025
- Report Date
- October 29, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244017818
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011110, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011110 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND PACKAGING IN THE MULTIVAC M12 ON 16-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5A01018 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 16-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 4M03928 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 20-DEC-2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5A02516 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 16-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKAGE EVENT ON (B)(6) 2025. THE LEAKAGE WAS IN THE QUICK RELEASE. THE INFUSION SET WAS IN USE FOR TWO DAYS. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224350 | QUICK SET | UNO QUICK-SET 80/9 SC1 MECA | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-386A | 6011110 | 05705244017818 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |