WOLVERINE CORONARY CUTTING BALLOON
Report
- Report Number
- 2124215-2025-75250
- Event Type
- Malfunction
- Date Received
- October 28, 2025
- Date of Event
- June 18, 2025
- Report Date
- October 28, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NWX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DAI KAWAUCHI, MD, KEI YUNOKI, MD, PHD, TOMOHIRO YOSHINO, MD, TAKEFUMI OKA, MD, PHD. CUTTING BALLOON BLADE FRACTURE IN SEVERE CALCIFIED LESIONS DURING PERCUTANEOUS CORONARY INTERVENTION 2025, VOL. 30, NO. 15 HTTPS://DOI/10.1016/J.JACCAS.2025.103773. B3 DATE OF EVENT: USED THE DATE THAT THE JOURNAL ARTICLE WAS PUBLISHED. DETAILED PRODUCT INFORMATION WAS NOT PROVIDED TO BSC. WE COMPLETED A GOOD FAITH EFFORT TO OBTAIN THE INFORMATION, BUT IT WAS UNABLE TO BE OBTAINED. BECAUSE THE LOT NUMBER IS UNKNOWN, WE ARE UNABLE TO PROVIDE THE UNIQUE IDENTIFIER (UDI) # AND OTHER SPECIFIC PRODUCT INFORMATION.
IT WAS REPORTED VIA LITERATURE THAT A WOLVERINE CB BECAME STUCK AND A BLADE FRACTURE OCCURRED. A CORONARY ANGIOGRAPHY (CAG) WAS PERFORMED AND SHOWED OCCLUDED LEFT ANTERIOR DESCENDING ARTERY (LAD), AND STENOTIC RIGHT CORONARY ARTERY (RCA) WITH LARGE, CALCIFIED LESION ON A PATIENT, HE WAS DIAGNOSED WITH CHRONIC CORONARY SYNDROME. A ROTATIONAL ATHERECTOMY WAS PERFORMED USING A .5 MM BURR. IVUS AFTERWARD SHOWED SEVERE STENOSIS OWING TO A LARGE, CALCIFIED NODULE; WIRE BIAS WAS NOT TOWARD THE LESION. A 2.5-10 MM WOLVERINE CB WAS INFLATED IN THE LESION IN A STEPWISE FASHION FROM 2-6 ATM IN 2 ATM INCREMENTS FOR 10-15 S AT EACH PRESSURE WHILE MOVING BACK AND FORTH. TEN INFLATIONS WERE PERFORMED. THE CB BECAME TEMPORARILY STUCK IN THE LESION AND SIGNIFICANT RESISTANCE WAS EXPERIENCED WHILE WITH DRAWING IT. VISUAL OBSERVATION OF THE CB CONFIRMED BLADE FRACTURE. NO CORONARY PERFORATION OR SEVERE DISSECTION WAS OBSERVED ON ANGIOGRAPHY OR IVUS. THE PATIENT WAS DISCHARGED. AT THE TIME OF PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR THE LAD 3MONTHS LATER, THERE WAS NO RESTENOSIS AT THE PCI SITE OF THE RCA, MONTHS LATER RCA RESTENOSIS WAS OBSERVED, A 4.0 -24 MMD RUG-ELUTING STENT WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2045883 | WOLVERINE CORONARY CUTTING BALLOON | CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING | NWX | BOSTON SCIENTIFIC CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Male |