WOLVERINE CORONARY CUTTING BALLOON
Report
- Report Number
- 2124215-2025-75139
- Event Type
- Malfunction
- Date Received
- October 28, 2025
- Date of Event
- June 18, 2025
- Report Date
- October 28, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NWX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DAI KAWAUCHI, MD, KEI YUNOKI, MD, PHD, TOMOHIRO YOSHINO, MD, TAKEFUMI OKA, MD, PHD. CUTTING BALLOON BLADE FRACTURE IN SEVERE CALCIFIED LESIONS DURING PERCUTANEOUS CORONARY INTERVENTION 2025, VOL. 30, NO. 15 HTTPS://DOI/10.1016/J.JACCAS.2025.103773. B3 DATE OF EVENT: USED THE DATE THAT THE JOURNAL ARTICLE WAS PUBLISHED. DETAILED PRODUCT INFORMATION WAS NOT PROVIDED TO BSC. WE COMPLETED A GOOD FAITH EFFORT TO OBTAIN THE INFORMATION, BUT IT WAS UNABLE TO BE OBTAINED. BECAUSE THE LOT NUMBER IS UNKNOWN, WE ARE UNABLE TO PROVIDE THE UNIQUE IDENTIFIER (UDI) # AND OTHER SPECIFIC PRODUCT INFORMATION.
IT WAS REPORTED VIA LITERATURE THAT A BALLOON WAS POORLY DILATED, AND A DOG-BONE SHAPE WAS OBSERVED, AND A BLADE FRACTURE OCCURRED. AN AVF. TRANSTHORACIC ECHO CARDIOGRAPHY REVEALED A MILD WALL MOTION ABNORMALITY IN THE INFERIOR LEFT VENTRICLE. HIGH-SENSITIVITY CARDIAC TROPONIN T CONCENTRATION WAS 2.290 NG ML (NORMAL, 0.014 NG ML). NON-ST-SEGMENT ELEVATION MYOCARDIAL INFARCTION WAS DIAGNOSED, AND EMERGENCY CORONARY ANGIOGRAPHY (CAG) WAS PERFORMED, WHICH SHOWED DISTAL TOTAL OCCLUSION OF THE RIGHT CORONARY ARTERY (RCA) AND SIGNIFICANT STENOSIS OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD). THE RCA OCCLUSION WAS PREDILATED USING A 2.0 COMPLIANT BALLOON, WHICH RESTORED FLOW; HOWEVER, BALLOON INDENTATION REMAINED. INTRAVASCULAR ULTRASOUND (IVUS) SHOWED SHORT LENGTH-CIRCUMFERENTIAL CALCIFICATION WITHIN THE CULPRIT LESION, WHICH HAD A MINIMUM LUMEN DIAMETER (MLD) OF 1.75 MM. THE ARTERIAL DIAMETER DISTAL TO THE LESION WAS 3.25 MM. A 2.75- 6-MM WOLVERINE CB WAS INFLATED IN A STEPWISE FASHION FROM 2-18 ATM IN 2-ATM INCREMENTS, HOLDING EACH PRESSURE FOR 10 TO 15 SECONDS. EIGHTEEN INFLATIONS WERE PERFORMED. THE CUTTING BALLOON WAS POORLY DILATED, AND A DOG-BONE SHAPE WAS OBSERVED; HOWEVER, THE INDENTATION DISAPPEARED AFTER A PROLONGED INFLATION AT 18ATM FOR 60 S. SIGNIFICANT RESISTANCE WAS EXPERIENCED WHILE PULLING THE CB. VISUAL OBSERVATION OF THE CB CONFIRMED BLADE FRACTURE. NO CORONARY PERFORATION OR SEVERE DISSECTION WAS SHOWN ON ANGIOGRAPHY. PATIENT WAS DISCHARGED ON POSTOPERATIVE DAY 14. ONE YEAR FOLLOW-UP CAG SHOWED NO RESTENOSIS AT THE PCI SITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1619361 | WOLVERINE CORONARY CUTTING BALLOON | CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING | NWX | BOSTON SCIENTIFIC CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Male |