WOLVERINE CORONARY CUTTING BALLOON
Report
- Report Number
- 2124215-2025-75167
- Event Type
- Malfunction
- Date Received
- October 28, 2025
- Date of Event
- June 18, 2025
- Report Date
- October 28, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NWX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DAI KAWAUCHI, MD, KEI YUNOKI, MD, PHD, TOMOHIRO YOSHINO, MD, TAKEFUMI OKA, MD, PHD. CUTTING BALLOON BLADE FRACTURE IN SEVERE CALCIFIED LESIONS DURING PERCUTANEOUS CORONARY INTERVENTION 2025, VOL. 30, NO. 15 HTTPS://DOI/10.1016/J.JACCAS.2025.103773 B3 DATE OF EVENT: USED THE DATE THAT THE JOURNAL ARTICLE WAS PUBLISHED. DETAILED PRODUCT INFORMATION WAS NOT PROVIDED TO BSC. WE COMPLETED A GOOD FAITH EFFORT TO OBTAIN THE INFORMATION, BUT IT WAS UNABLE TO BE OBTAINED. BECAUSE THE LOT NUMBER IS UNKNOWN, WE ARE UNABLE TO PROVIDE THE UNIQUE IDENTIFIER (UDI) # AND OTHER SPECIFIC PRODUCT INFORMATION.
IT WAS REPORTED VIA LITERATURE THAT A BLADE FRACTURE AND PARTIAL DROPOUT OCCURRED. A PATIENT WAS UNDERWENT FOR RIGHT CORONARY ARTERY (RCA) PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR CHRONIC CORONARY SYNDROME. CARDIOVASCULAR AND RESPIRATORY PHYSICAL FINDINGS WERE NORMAL, AND NEUROLOGIC EXAMINATION REVEALED MILD LEFT HEMIPARESIS. IVUS WAS USED AND AS IT DID NOT PASS THROUGH THE LESION, ROTATIONAL ATHERECTOMY WAS PERFORMED USING A 1.25-MM BURR. IVUS AFTERWARD SHOWED A SHORT SEGMENT CIRCUMFERENTIAL SEVERE CALCIFICATION PROXIMAL TO THE TARGET LESION WITH A MLD OF 1.95 MM. ARTERIAL DIAMETER DISTAL TO THE CALCIFICATION WAS 3.12 MM. AFTER PERFORMING ORBITAL ATHERECTOMY OF THE CALCIFICATION, IT WAS DILATED USING A 2.5- 6-MM WOLVERINE CB, A DOG BONE-SHAPE WAS OBSERVED. THE CB WAS INFLATED IN A STEPWISE FASHION FROM 2 TO 12 ATMS IN INCREMENTS OF 2 ATMS. EIGHTEEN INFLATIONS WERE PERFORMED. AFTER A PROLONGED 45-S INFLATION AT 12 ATM, THE BALLOON INDENTATION ALMOST DISAPPEARED. THE CB CATHETER WAS PULLED OUT WITHOUT RESISTANCE. BLADE FRACTURE AND PARTIAL DROPOUT WERE OBSERVED. NO CORONARY PERFORATION OR SEVERE DISSECTION WAS SHOWN ON ANGIOGRAPHY. THE PATIENT SYMPTOMS RESOLVED, AND FOLLOW-UP CAG 6 MONTHS LATER SHOWED NO RESTENOSIS, NO RESIDUAL FRACTURED CB BLADE WAS OBSERVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1619359 | WOLVERINE CORONARY CUTTING BALLOON | CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING | NWX | BOSTON SCIENTIFIC CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Male |