FDA Adverse Event Injury Summary report: N

INSET GUARD

MDR report key: 23401373 · Received October 28, 2025

Report

Report Number
8021545-2025-02927
Event Type
Injury
Date Received
October 28, 2025
Date of Event
September 21, 2025
Report Date
November 4, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018877
PMA / PMN Number
K173879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (8021545-2025-02927), WAS SUBMITTED ON 28-OCT-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 03-JUL-2025. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6013616. IN QUESTION WAS MANUFACTURED AT THE OSTED SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 04-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH" ,"LOT NUMBER" CRITERIA EQUAL 6013616. THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE 6013616 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) [80] APPENDIX 1 BATCHCARD FOR PRODUCTION OF PACKAGING ROOM ON 03-JUL-2025 WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PRODUCT WAS RETURNED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT WAS HOSPITALIZED DUE TO HYPERGLYCEMIA ON (B)(6) 2025. BLOOD GLUCOSE LEVEL AT THE TIME OF EVENT WAS HIGH AND PATIENT WAS TREATED WITH INSULIN VIA INTRAVENOUS (IV) DRIP. THE PATIENT ALSO HAD SYMPTOMS OF FEELING SICK/ UNWELL, TIREDNESS/ WEAKNESS AND VOMITING. PATIENT WAS ALSO FOUND POSITIVE FOR HIGH KETONES. LENGTH OF HOSPITALIZATION WAS GREATER THAN 24 HOURS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2293006 INSET GUARD UNO MIO ADV. GREY 60/6 SC1 10-PK INT FPA UNOMEDICAL A/S MMT-242A 6013616 05705244018877

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H