FDA Adverse Event Malfunction Summary report: N

ULTIPACE¿ LEAD, 65CM MRI

MDR report key: 23400382 · Received October 28, 2025

Report

Report Number
2017865-2025-1003423
Event Type
Malfunction
Date Received
October 28, 2025
Report Date
December 9, 2025
Manufacturer
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Product Code
NVN
UDI-DI
05415067046321
PMA / PMN Number
P960013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FURTHER INFORMATION WAS REQUESTED BUT NOT RECEIVED.

Additional Manufacturer Narrative · 0

THE REPORTED EVENTS OF FAILURE TO CAPTURE AND NOISE WERE NOT CONFIRMED. A COMPLETE LEAD WAS RETURNED FOR ANALYSIS. ELECTRICAL TESTING DID NOT FIND ANY INDICATION OF CONDUCTOR FRACTURES OR INTERNAL SHORTS. VISUAL AND X-RAY INSPECTION OF THE LEAD DID NOT FIND ANY ANOMALIES.

Additional Manufacturer Narrative · 0

CORRECTION: SECTION G3. THE CORRECT INITIAL AWARENESS DATE SHOULD HAVE BEEN (B)(6) 2025, RATHER THAN (B)(6) 2025 IN 2017865-2025-1003423.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT HAD PRESENTED FOR A FOLLOW-UP VISIT AT THE CLINIC. UPON INTERROGATION, IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD HAD EXHIBITED LOSS OF CAPTURE. AN X-RAY REVEALED THAT THE RA LEAD HAD DISLODGED, AND IT WAS SUBSEQUENTLY EXPLANTED AND REPLACED. DURING THE RIGHT VENTRICULAR (RV) LEAD REVISION FOR AN UPGRADE TO LEFT BUNDLE PACING, IT WAS OBSERVED THAT THE FIRST NEW RV LEAD HAD FAILED TO CAPTURE AND EXHIBITED NOISE REVERSION WHEN CONNECTED TO THE CHRONIC PACEMAKER. THE ISSUE PERSISTED EVEN AFTER THE LEAD WAS CONNECTED TO A NEW PACEMAKER, BUT IT WAS RESOLVED FOLLOWING APPROPRIATE PROGRAMMING OF THE RV LEAD. HOWEVER, THE FIRST NEW RV LEAD WAS NOT USED. THE PHYSICIAN ELECTED TO IMPLANT A SECOND NEW RV LEAD AND SUCCESSFULLY COMPLETED THE PROCEDURE. THE PATIENT REMAINED STABLE THROUGHOUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2735516 ULTIPACE¿ LEAD, 65CM MRI NO MATCH NVN ST. JUDE MEDICAL, INC.(CRM-SYLMAR) LPA1231/65 P000231145 05415067046321

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male ASSURITY MRI PACEMAKER.| ULTIPACE LEAD.