QUICK SET
Report
- Report Number
- 3003442380-2025-15326
- Event Type
- Malfunction
- Date Received
- October 27, 2025
- Date of Event
- September 29, 2025
- Report Date
- October 31, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244017573
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: ARGENTINA.
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011897 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011897 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND PACKAGING IN THE MULTIVAC M12 ON 02-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5B03932 WAS MANUFACTURED ACCORDING TO THE WI VERSION 28 AND ASSEMBLED IN THE QUICKSET LINE, ON 02-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5B03933 WAS MANUFACTURED ACCORDING TO THE WI VERSION 28 AND ASSEMBLED IN THE QUICKSET LINE, ON 02-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5B02969 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 21-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5B03918 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 26-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5C00349 WAS MANUFACTURED ACCORDING TO THE WI VERSION 28 AND ASSEMBLED IN THE QUICKSET LINE, ON 03-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: ASSEMBLY OF THE LOT 5B04161 WAS MANUFACTURED ACCORDING TO THE WI VERSION 27 AND ASSEMBLED IN THE QUICKSET LINE, ON 21-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B04587 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE LINE MP04 AND MP08, ON 28-FEB-2025, WITH A TOTAL OF 120,000 UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B03916 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE LINE MP04 AND MP08, ON 26-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B03915 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE LINE MP04 AND MP08, ON 25-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B02965 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE LINE MP04 AND MP08, ON 19-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN ARGENTINA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET TUBING DETACHMENT EVENT ON (B)(6) 2025. THE SITE OF DETACHMENT WAS QUICK RELEASE. THE INSERTION SITE WAS ABDOMEN. THE INFUSION SET WAS IN USE FOR FEW HOURS. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19409 | QUICK SET | UNO QUICK-SET 60/9 SC1 MECA | FPA | UNOMEDICAL A/S | MMT-397A | 6011897 | 05705244017573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |