FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD

MDR report key: 23386524 · Received October 24, 2025

Report

Report Number
1710034-2025-01658
Event Type
Malfunction
Date Received
October 24, 2025
Date of Event
April 17, 2024
Report Date
October 27, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903818235
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: REPORTING AWARENESS DATE (DATE RECEIVED BE THE MFG) IS THE DATE OF THE INVESTIGATION CLOSURE. DATE HAS BEEN CORRECTED.

Additional Manufacturer Narrative · 0

OUR QUALITY ENGINEER INSPECTED THE SAMPLE FOR EVALUATION. YOUR REPORT OF NEEDLE RETRACTION FAILURE WAS CONFIRMED, AND THE CAUSE WAS DETERMINED TO BE MANUFACTURING RELATED. ONE 22GA INSYTE AUTOGUARD UNIT FROM LOT NUMBER: 3251086 WAS PROVIDED FOR INVESTIGATION. THE NEEDLE WAS RECEIVED FULLY EXTENDED FROM THE SHIELD. ADHESIVE WAS FOUND BETWEEN THE NEEDLE HUB AND THE GRIP, WHICH PREVENTED NEEDLE RETRACTION. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCE'S ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. HOWEVER, A TREND WAS IDENTIFIED FOR THE REPORTED FAILURE AND CORRECTIVE ACTIONS HAVE BEEN IMPLEMENTED TO INVESTIGATE THIS TYPE OF INCIDENT AND IDENTITY THE ROOT CAUSE. THE APPROPRIATE MANUFACTURING PERSONNEL WERE NOTIFIED OF THIS COMPLAINT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

HARD NEEDLES CANNOT BE RECYCLED NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2217192 INSYTE AUTOGUARD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 3251086 00382903818235

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown