FDA Adverse Event Injury Summary report: N

INSET GUARD

MDR report key: 23378352 · Received October 24, 2025

Report

Report Number
8021545-2025-02888
Event Type
Injury
Date Received
October 24, 2025
Date of Event
September 27, 2025
Report Date
November 20, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018877
PMA / PMN Number
K173879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (8021545-2025-02888), WAS SUBMITTED ON 24-OCT-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 28-APR-2025. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6012791 IN QUESTION WAS MANUFACTURED AT THE OSTED SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 19-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6012791". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3 SO NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012791 WAS MANUFACTURED ACCORDING TO APPENDIX 1 BATCHCARD FOR PRODUCTION OF PACKAGING ROOM ON 28-APR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO SAMPLE WAS PROVIDED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT WENT TO HOSPITAL WITH HIGH BLOOD GLUCOSE DUE TO INFUSION SET CAME OFF ON (B)(6) 2025. THE INSERTION SITE WAS RIGHT AND LEFT ABDOMEN. THE INFUSION SET WAS IN USE FOR 2 DAYS. THE PATIENT GOT TREATED WITH INSULIN VIA INTRAVENOUS (IV) DRIP. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
603165 INSET GUARD UNO MIO ADV. GREY 60/6 SC1 10-PK INT FPA UNOMEDICAL A/S MMT-242A 6012791 05705244018877

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention