FDA Adverse Event Injury Summary report: N

LINEAR? ST

MDR report key: 23376836 · Received October 24, 2025

Report

Report Number
3006630150-2025-09517
Event Type
Injury
Date Received
October 24, 2025
Date of Event
September 22, 2025
Report Date
October 24, 2025
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729767725
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK B3: EXACT DATE UNKNOWN, EVENT OCCURRED A WEEK AGO FROM THE DATE THE MANUFACTURER WAS MADE AWARE. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS, UPN: M365SC2218500, MODEL: SC-2218-50, SERIAL: (B)(6) BATCH: 7162858, UDI: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT EXPERIENCED A RIB STIMULATION DUE TO LEAD MIGRATION THAT WAS CONFIRMED VIA X-RAY. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WILL NOT BE RETURNED PER HOSPITAL POLICY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
229007 LINEAR? ST STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-2218-50 7156804 08714729767725

Patients

Seq Age Sex Outcome Treatment
1 38 YR Male Required Intervention