ARCHITECT TOTAL PSA
Report
- Report Number
- 3008344661-2025-00146
- Event Type
- Malfunction
- Date Received
- October 23, 2025
- Date of Event
- October 14, 2025
- Report Date
- December 10, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MTF
- UDI-DI
- 00380740159849
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORTED THE COMPLAINT ISSUE WITHOUT INDICATION OF ANY ADDITIONAL ISSUES. A SEARCH FOR SIMILAR COMPLAINTS DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR LOT 73579FZ01. A REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT TOTAL PSA ASSAY DID NOT IDENTIFY ANY TRENDS RELATED TO THE COMPLAINT ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES OR DEVIATIONS WITH LOT NUMBER 73579FZ01AND THE COMPLAINT ISSUE. IN-HOUSE PERFORMANCE TESTING WAS PERFORMED UTILIZING RETAINED REAGENT KIT OF THE COMPLAINT LOT. ALL TESTING PASSED INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER¿S ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT TOTAL PSA REAGENT, LOT NUMBER73579FZ01.
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K70-74 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 6C06, WITH PMA NUMBER P910007. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE RESULTS WERE NOT REPORTED OUT OF THE LAB. THE TOTAL PSA UNIT IS NG/ML, WITH A REFERENCE RANGE OF 0-4 NG/ML. THE SPECIFIC DATA ARE AS FOLLOWS: SID 2 INITIAL RESULT 0.002 NG/ML, REPEATED 6.087 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE RESULTS WERE NOT REPORTED OUT OF THE LAB. THE TOTAL PSA UNIT IS NG/ML, WITH A REFERENCE RANGE OF 0-4 NG/ML. THE SPECIFIC DATA ARE AS FOLLOWS: SID 2 INITIAL RESULT 0.002 NG/ML, REPEATED 6.087 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1931242 | ARCHITECT TOTAL PSA | TOTAL, PROSTATE SPECIFIC AG (NONCOMPLEXED/COMPLEXED) FOR DET OF PROSTATE CANCER | MTF | ABBOTT IRELAND DIAGNOSTICS DIVISION | 73579FZ01 | 00380740159849 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2000SR INST, 03M74-02, (B)(6) |