FDA Adverse Event Malfunction Summary report: N

ARCHITECT TOTAL PSA

MDR report key: 23366650 · Received October 23, 2025

Report

Report Number
3008344661-2025-00146
Event Type
Malfunction
Date Received
October 23, 2025
Date of Event
October 14, 2025
Report Date
December 10, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MTF
UDI-DI
00380740159849
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORTED THE COMPLAINT ISSUE WITHOUT INDICATION OF ANY ADDITIONAL ISSUES. A SEARCH FOR SIMILAR COMPLAINTS DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR LOT 73579FZ01. A REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT TOTAL PSA ASSAY DID NOT IDENTIFY ANY TRENDS RELATED TO THE COMPLAINT ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES OR DEVIATIONS WITH LOT NUMBER 73579FZ01AND THE COMPLAINT ISSUE. IN-HOUSE PERFORMANCE TESTING WAS PERFORMED UTILIZING RETAINED REAGENT KIT OF THE COMPLAINT LOT. ALL TESTING PASSED INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER¿S ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT TOTAL PSA REAGENT, LOT NUMBER73579FZ01.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K70-74 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 6C06, WITH PMA NUMBER P910007. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE RESULTS WERE NOT REPORTED OUT OF THE LAB. THE TOTAL PSA UNIT IS NG/ML, WITH A REFERENCE RANGE OF 0-4 NG/ML. THE SPECIFIC DATA ARE AS FOLLOWS: SID 2 INITIAL RESULT 0.002 NG/ML, REPEATED 6.087 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE RESULTS WERE NOT REPORTED OUT OF THE LAB. THE TOTAL PSA UNIT IS NG/ML, WITH A REFERENCE RANGE OF 0-4 NG/ML. THE SPECIFIC DATA ARE AS FOLLOWS: SID 2 INITIAL RESULT 0.002 NG/ML, REPEATED 6.087 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1931242 ARCHITECT TOTAL PSA TOTAL, PROSTATE SPECIFIC AG (NONCOMPLEXED/COMPLEXED) FOR DET OF PROSTATE CANCER MTF ABBOTT IRELAND DIAGNOSTICS DIVISION 73579FZ01 00380740159849

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ARC I2000SR INST, 03M74-02, (B)(6)