FDA Adverse Event Injury Summary report: N

COMFORT

MDR report key: 23364497 · Received October 23, 2025

Report

Report Number
3003442380-2025-15190
Event Type
Injury
Date Received
October 23, 2025
Date of Event
September 19, 2025
Report Date
October 24, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244019614
PMA / PMN Number
K162812
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PATIENT COUNTRY: CANADA. PATIENT CITY: (B)(6).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4) THE BATCH 6010051, IN QUESTION WAS MANUFACTURED AT THE: REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 17-OCT-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010051". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010051 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 20 AND MANUFACTURED IN THE MULTIVAC 14 ON 11-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THE SUB-ASSEMBLY: THE CANNULA LOT 4K02160 WAS MANUFACTURED ACCORDING TO THE WI VERSION 36 AND MANUFACTURED ON 09-OCT-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. GLUING OF TUBING OF THE LOT 4K05855 MANUFACTURED IN THE MACHINE LC01, ON 10-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED, THEREFORE, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH: WI GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3: 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2: 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST FOR THE CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. ALL TEST RESULTS WERE WITHIN SPECIFICATIONS AS PER THE TEST REPORT (B)(4). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO DEFECT ON TEST OF REFERENCE SAMPLES., NO NC RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED ON (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE BLOOD GLUCOSE LEVEL WAS OVER 30 MMOL/L AT THE TIME OF EVENT. THE PATIENT REPORTED CONFUSION AT THE TIME OF EVENT. THE PATIENT WAS IN THE HOSPITAL FOR MORE THAN TWENTY-FOUR HOURS. THE PATIENT GOT TREATED WITH INTRAVENOUS INSULIN. THE PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERY SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139923 COMFORT UNO COMFORT SHORT 110/13 SC1 MINI FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-382A 6010051 05705244019614

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R