FDA Adverse Event Malfunction Summary report: N

ACTIVFLO ROUTINE I-TAPED- BLUE 1000

MDR report key: 23362929 · Received October 22, 2025

Report

Report Number
1419341-2025-00036
Event Type
Malfunction
Date Received
October 22, 2025
Date of Event
August 21, 2025
Report Date
November 5, 2025
Manufacturer
XIAMEN VOKE MOLD & PLASTIC ENGINEER
Product Code
IDZ
UDI-DI
20849832022378
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

RETAIN SAMPLES WERE NOT AVAILABLE FOR FURTHER TESTING. THE ROOT CAUSE FOR "THE PLASTIC SEEMS QUITE FLIMSY, AND STAFF HAVE HAD ISSUES WITH THEM CLOSING PROPERLY" REPORTED BY THE CUSTOMER COULD NOT BE UNEQUIVOCALLY DETERMINED FROM THE INFORMATION AVAILABLE. ALTHOUGH THE INFORMATION AVAILABLE INDICATES THAT NO ADVERSE CONSEQUENCE(S) TO A PATIENT(S) HAS BEEN REPORTED TO THE MANUFACTURER, THE CIRCUMSTANCES INVOLVED IN THIS COMPLAINT HAVE PREVIOUSLY RESULTED IN SERIOUS INJURY. IF ADDITIONAL INFORMATION IS RECEIVED AN ADDITIONAL FOLLOW UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE CUSTOMER REPORTED ACTIVFLO ROUTINE I-TAPED- BLUE 1000, P/N 39LC-500-2-L, LOT 20250307 SEEMED QUITE FLIMSY, AND STAFF HAVE HAD ISSUES WITH THEM CLOSING PROPERLY. NO ADVERSE EVENTS WERE REPORTED. TRENDING ANALYSIS FROM 15 OCT 2024 TO 15 OCT 2025 REPORTED NO (0) OTHER RELATED OCCURRENCES OF THIS ISSUE FOR THIS PRODUCT LOT. AS MANUFACTURER INVESTIGATION OF THIS COMPLAINT BY LEICA BIOSYSTEMS IS IN PROGRESS, THE ROOT CAUSE OF THIS EVENT HAS NOT YET BEEN DETERMINED.

Description of Event or Problem · 0

ON 25 SEPTEMBER 2025, LEICA BIOSYSTEMS RECEIVED A COMPLAINT DETAILING "THE PLASTIC SEEMS QUITE FLIMSY, AND STAFF HAVE HAD ISSUES WITH THEM CLOSING PROPERLY." ON 18 OCTOBER 2025, THE LEICA ACCOUNT MANAGER ADVISED THE LEICA ASSOCIATE, QUALITY ASSURANCE THAT THE AWARENESS DATE FOR THIS EVENT WAS 21 AUGUST 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
229702 ACTIVFLO ROUTINE I-TAPED- BLUE 1000 CASSETTES, TISSUE IDZ XIAMEN VOKE MOLD & PLASTIC ENGINEER ACTIVFLO ROUTINE I-TAPED- BLUE 1000 20250307 20849832022378

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown