FDA Adverse Event Injury Summary report: N

RESONANCE STENT SET

MDR report key: 23356155 · Received October 22, 2025

Report

Report Number
3001845648-2025-00532
Event Type
Injury
Date Received
October 22, 2025
Date of Event
January 12, 2022
Report Date
December 4, 2025
Manufacturer
COOK IRELAND LTD
Product Code
FAD
PMA / PMN Number
K063742
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS STILL PENDING, A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: ABNORMAL USE COMPLAINTS ARE CONSIDERED TO BE BEYOND ANY FURTHER REASONABLE MEANS OF INTERFACE-RELATED RISK CONTROL BY THE MANUFACTURER. IT IS UNKNOWN HOW THE DEVICE WILL FUNCTION OUTSIDE OF ITS INTENDED USE. TRENDING WILL MONITOR IF ANY FUTURE INVESTIGATION IS REQUIRED THE DEVICE EVALUATION OF RMS-060024-R DEVICE OF UNKNOWN LOT AND CATALOG NUMBER COULD NOT BE COMPLETED AS THE DEVICE OR PHOTOGRAPHIC EVIDENCE OF THE DEVICE WAS NOT RETURNED FOR EVALUATION. THIS FILE WAS CREATED IN RESPONSE TO THE ATTACHED PMCF STUDY AND RELATES TO STENT MIGRATION MANUFACTURING RECORDS: PRIOR TO DISTRIBUTION ALL DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL INSPECTION TO ENSURE DEVICE INTEGRITY. MANUFACTURING RECORDS REVIEW COULD NOT BE COMPLETED AS THE LOT NUMBER IS UNKNOWN. INSTRUCTIONS FOR USE AND/LABEL INSTRUCTIONS FOR USE (IFU0020) STATES THE FOLLOWING: ¿INTENDED USE: USED FOR TEMPORARY STENTING OF THE URETER IN ADULT PATIENTS WITH EXTRINSIC URETERAL OBSTRUCTION.¿. THERE IS EVIDENCE TO SUGGEST THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE (IFU0020). IT IS ALSO KNOWN FROM THE AVAILABLE INFORMATION THAT THE INCORRECT WIRE GUIDE WAS USED. AS PER MANAGEMENT OF COMPLAINTS PROCEDURE WHERE INFORMATION IS REVIEWED AND CATEGORIZED AS USE RELATED/OFF LABEL USE AND WHERE NO ADVERSE EVENT IS IDENTIFIED THEN NO COMPLAINT IS REQUIRED. THE EVENT IS DOCUMENTED IN THE PMS TRACKER AND REVIEWED AS PART OF POST MARKET SURVEILLANCE. IMAGE REVIEW: AN IMAGE WAS NOT RETURNED FOR EVALUATION. ROOT CAUSE ANALYSIS: DEFINITIVE ROOT CAUSE WAS ESTABLISHED. THE USER HAS NOT COMPLIED WITH THE REQUIREMENTS OF THE IFU WITH RESPECT TO THE INTENDED USE, INDICATIONS FOR USE OR CONTRAINDICATIONS OF THE DEVICE. IT IS KNOWN FROM THE AVAILABLE INFORMATION THAT THE STENT WAS PLACED TO TREAT AN INTRINSIC OBSTRUCTION. THE MIGRATION OF THE STENT WAS MOST LIKELY CAUSE BY THE ABNORMAL USE. CONFIRMATION OF COMPLAINT: COMPLAINT IS CONFIRMED BASED ON CUSTOMER AND/OR REP TESTIMONY. CORRECTIVE ACTION/ CORRECTION: COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR SIMILAR EVENT. SUMMARY: ACCORDING TO PMCF STUDY THERE WAS 01 CASE OF STENT MIGRATION/ DISPLACEMENT REPORTED. 01 CONFIRMED QUANTITY OF 01 DEVICE, CONFIRMED USED. THE STENT WAS REMOVED. DEFINITIVE ROOT CAUSE WAS ESTABLISHED. THE USER HAS NOT COMPLIED WITH THE REQUIREMENTS OF THE IFU WITH RESPECT TO THE INTENDED USE, INDICATIONS FOR USE OR CONTRAINDICATIONS OF THE DEVICE. IT IS KNOWN FROM THE AVAILABLE INFORMATION THAT THE STENT WAS PLACED TO TREAT AN INTRINSIC OBSTRUCTION. THE MIGRATION OF THE STENT WAS MOST LIKELY CAUSE BY THE ABNORMAL USE.

Description of Event or Problem · 0

AS REPORTED TO CUSTOMER RELATIONS VIA CRI OBSERVATION STUDY COMPLAINT FORM: ON (B)(6) 2020 PATIENT UNDERWENT STENT PLACEMENT TO THE RIGHT URETER VIA URETEROSCOPY DUE TO INTRINSIC OBSTRUCTION/STRICTURE. PROCEDURE WAS DONE UNDER FLUOROSCOPY. SITE INDICATED THE PROCEDURE WAS SUCCESSFUL. ON (B)(6) 2021, 138 DAYS INDWELL TIME THE DEVICE WAS REMOVED DUE TO SCHEDULED REPLACEMENT. ON (B)(6) 2022, 336 DAYS INDWELL TIME THE DEVICE WAS ATTEMPTED TO BE REMOVED. SITE INDICATED DEVICE DEFICIENCY DUE TO STENT DISLODGEMENT/MIGRATION. THE SITE STATED: ¿THIS WAS A SCHEDULED REPLACEMENT. HOWEVER, STENT WAS ULTIMATELY UNABLE TO BE REMOVED. DURING THE OPERATION, PER OPERATIVE NOTE ON (B)(6) 2022, FINDING WAS "PROXIMAL MIGRATION OF RESONANCE STENT INTO DISTAL URETER, DESPITE MULTIPLE ATTEMPTS, UNABLE TO BASKET AND REMOVE." ON THE SAME DAY THE SITE REPORTED ADVERSE EVENT OF HYDRONEPHROSIS. TREATMENT INCLUDED STENT REMOVAL. ON (B)(6) 2025 A SECOND ATTEMPT WAS MADE AND SUCCESSFUL REMOVAL WITH DEVICE REPLACEMENT. PER THE SITE: ¿STENT WAS UNABLE TO BE REMOVED ON (B)(6) 2022. IT WAS EXCHANGED ON (B)(6) 2022 AFTER A SECOND ATTEMPT¿. SITE HAS RELATED THE EVENT TO THE RESONANCE METALLIC URETERAL STENT. PER THE OPERATIVE NOTE ON (B)(6) 2022, "THE KIDNEY DEMONSTRATED SEVERE HYDRONEPHROSIS WITH FILLING DEFECTS IN THE RENAL PELVIS CORRESPONDING TO THE EXISTING STENT." THE EVENT WAS NOT RELATED TO ANY COOK ANCILLARY DEVICE, STUDY PROCEDURE AND NO OTHER CONDITION OR CIRCUMSTANCE HAS CAUSED OR CONTRIBUTED TO THIS EVENT. THE SITE REPORTED A DEVICE DEFICIENCY PER THE OPERATIVE NOTE: "PROXIMAL MIGRATION OF RESONANCE STENT INTO DISTAL URETER, DESPITE MULTIPLE ATTEMPTS, UNABLE TO BASKET AND REMOVE." THE SITE STATED THE STENT WAS REPLACED ON (B)(6) 2022 WHICH CONCLUDED THE PATIENT¿S ENROLLMENT IN THE STUDY ADDITIONAL QUESTIONS ASKED 03-OCT-2025. 1. ARE ANY IMAGES AVAILABLE FOR REVIEW? N/A, YES, NO - ANS. NO. 2. WHAT WAS THE TARGET LOCATION FOR THE STENT? ANS. RIGHT URETER, WITH THE PROXIMAL CURL IN THE RENAL PELVIS AND DISTAL CURL IN THE BLADDER 3. DID ANYTHING HAVE TO BE REMOVED FROM THE PATIENT? N/A, YES, NO - ANS. YES. 4. IF YES, PLEASE SPECIFY: A. WHAT PART OF THE BODY THE DEVICE WAS REMOVED FROM? ANS. RIGHT URETER AND RENAL PELVIS. B. WHAT DEVICE WAS REMOVED? ANS. RIGHT URETER AND RENAL PELVIS. C. WHAT INSTRUMENT WAS USED TO REMOVE IT? ANS. RIGHT URETER AND RENAL PELVIS. 5. PLEASE SPECIFY THE STORAGE CONDITIONS OF THE DEVICE AT THE FACILITY, PARTICULARLY THOSE RELATING TO LIGHT AND TEMPERATURE ANS. UNSPECIFIED/UNKNOWN. 6. WHY WAS THE STENT REMOVED? N/A, EXCHANGE, OTHER (IF OTHER, PLEASE DETAIL WHY THE STENT WAS REMOVED) ANS. THIS WAS A SCHEDULED REPLACEMENT. HOWEVER, STENT WAS ULTIMATELY UNABLE TO BE REMOVED. DURING THE OPERATION, PER OPERATIVE NOTE ON (B)(6) 2022, FINDING WAS "PROXIMAL MIGRATION OF RESONANCE STENT INTO DISTAL URETER, DESPITE MULTIPLE ATTEMPTS, UNABLE TO BASKET AND REMOVE." STENT WAS SUCCESSFULLY REMOVED ON (B)(6) 2022. 7. IF THE STENT WAS REMOVED WHAT WAS USED TO COMPLETE THE PROCEDURE? ANS. NEW 6FR BY 24CM RESONANCE STENT PLACED UNDER ENDOSCOPIC AND FLUOROSCOPIC GUIDANCE. 8. WHAT WAS THE LENGTH OF THE INDEWLL TIME? ANS. 337 DAYS 9. WHAT DISEASE MODE WAS THE PHYSICIAN TRYING TO TREAT? ANS. URETERAL OBSTRUCTION/PSEUDO-OBSTRUCTION RELATED TO REDUNDANT RENAL PELVIS CAUSING IMPAIRED DRAINAGE IN A SOLITARY RIGHT KIDNEY. 10. WHAT TYPE AND SIZE WIRE GUIDE WAS USED WITH THE DEVICE? ANS. 0.035 SENSOR WIRE AND A SUPERSTIFF WIRE. 11. DID THE DEVICE COME WITH A PIGTAIL STRAIGHTENER? N/A, YES, NO (IF YES, WAS THE PIGTAIL STRAIGHTENER USED? N/A, YES, NO) ANS. UNSPECIFIED/UNKNOWN. 12. DID THE USER ATTEMPT TO ATTACH THE STENT TO THE INSERTER BEFORE OR AFTER WIRE GUIDE INSERTION? N/A, BEFORE, AFTER ANS. UNSPECIFIED/UNKNOWN. 13. DID THE USER ATTEMPT TO ATTACH THE TAPERED END OF THE STENT TO THE INSERTER? N/A, YES, NO ANS. UNSPECIFIED/UNKNOWN. 14. WAS THE DEVICE ASSEMBLED OUTSIDE OF THE BODY? N/A, YES, NO ANS. UNSPECIFIED/UNKNOWN. 15. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES AS A RESULT OF THIS EVENT? N/A, YES, NO - ANS. NO 16. WHAT INTERVENTION (IF ANY) WAS REQUIRED? ANS. REMOVAL OF OLD STENT AND PLACEMENT OF NEW STENT. 17. WAS THE SECONDARY INTERVENTION PERFORMED DURING THE SAME PROCEDURE AS THE DEVICE FAILURE OR WAS IT SCHEDULED FOR ANOTHER DAY? ANS. N/A, SAME PROCEDURE, ANOTHER DAY -ANOTHER DAY; ATTEMPTED REMOVAL (B)(6) 2022 WHICH WAS UNSUCCESSFUL, SECOND ATTEMPT (B)(6) 2022 SUCCESSFUL. 18. WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DEVICE PRIOR TO RETURN (E.G., KINK)? N/A, YES, NO (IF YES, PLEASE SPECIFY WHAT WAS OBSERVED AND WHERE ON THE DEVICE IT WAS OBSERVED) ANS. UNSPECIFIED/UNKNOWN. 19. HOW OFTEN WAS THE STENT CHECKED DURING THE IN-DWELLING TIME? (WHAT METHOD WAS USED?) ANS. 1 ULTRASOUND WAS PERFORMED. ¿RIGHT: THE RIGHT KIDNEY MEASURES 15.1 X 7 X 6.1 CM AND DEMONSTRATES NORMAL CORTICAL ECHOGENICITY AND NORMAL CONTOUR. THERE IS MILD PELVIECTASIS. THERE IS NO FOCAL RENAL MASS LESION OR SHADOWING STONE. A DOUBLE-J URETERIC STENT EXTENDS FROM THE RENAL PELVIS INTO THE BLADDER LUMEN. LEFT: THE LEFT KIDNEY IS SURGICALLY ABSENT. THE URINARY BLADDER APPEARS UNREMARKABLE FOR DEGREE OF DISTENSION. IMPRESSION: 1. A DOUBLE-J RIGHT URETERIC STENT EXTENDS FROM THE RENAL PELVIS INTO THE BLADDER LUMEN. EVIDENCE OF MILD RIGHT PELVIECTASIS BUT NO FRANK HYDRONEPHROSIS. 2. COMPENSATORY HYPERTROPHY OF THE RIGHT KIDNEY AND SURGICAL ABSENCE OF THE LEFT KIDNEY.¿ 20. WAS THE PATIENT USING CALCIUM SUPPLEMENTATION? ANS. NO. 21. WAS FORCE REQUIRED TO REMOVE THE STENT? ANS. NO. 22. WAS ENCRUSTATION EVIDENT ON THE STENT? ANS. NO. 23. "ACCORDING TO THE RESONANCE METALLIC URETERAL STENT IFU, THE DEVICE IS INDICATED FOR UP TO O12 MONTHS OF INDWELL TIME. THE COMPLAINT REPORT STATES THAT THE DEVICE WAS REMOVED AFTER 138 DAYS DUE TO 'SCHEDULED REPLACEMENT.' CAN YOU CLARIFY WHY THE DEVICE WAS SCHEDULED FOR REMOVAL AT 138 DAYS, RATHER THAN THE MAXIMUM 12 MONTH INDWELL PERIOD AS PERMITTED BY THE IFU? WAS THERE A CLINICAL OR PROCEDURAL REASON FOR THIS EARLIER REMOVAL?" ANS .UNCLEAR, BUT IN THE INDICATIONS FOR THE PROCEDURE IN THE OP NOTE ON (B)(6) 2021, ¿58 YEAR OLD MALE WITH SOLITARY RIGHT KIDNEY, INITIALLY PRESENTED WITH RUPTURE OF RIGHT RENAL COLLECTING SYSTEM WITH SIGNIFICANT RETROPERITONEAL URINE EXTRAVASATION WHO WAS MANAGED WITH A RIGHT NEPHROSTOMY TUBE AND SUBSEQUENTLY UNDERWENT DIAGNOSTIC ANTEGRADE/RETROGRADE. ON 5/11 UNDERWENT ROBOTIC RIGHT PYELOLASTY. STENT WAS REMOVED AFTER 6 WEEKS. SUBSEQUENTLY HAS HAD WORSENING RIGHT FLANK PAIN. LASIX RENAL SCAN SHOWS OBSTRUCTION. CT WITH HYDRO. AKI WITH CR TO 1.5. HE UNDERWENT URETEROSCOPY WHICH NOTED A PATENT ANASTOMOSIS WITH REDUNDANT RENAL PELVIS FLOPPING OVER THE URETER CAUSING A PSEUDO-OBSTRUCTION. A NEPHROSTOMY TUBE WAS PLACED FOR TEMPORARY DRAINAGE OF HIS SOLITARY RIGHT KIDNEY. GIVEN WIDE OPEN ANASTOMOSIS OPTION TO PROCEED WITH REPEAT ANTEGRADE NEPHROSTOGRAM. HOWEVER, STILL DID NOT SHOW PASSAGE OF CONTRAST. HE UNDERWENT A REDUCTION PYELOPLASTY GIVEN THE LARGE REDUNDANT RENAL PELVIS. INTRAOP, WAS NOTED TO HAVE WIDE OPEN PATENT ANASTOMOSIS AS WELL AS PASSAGE OF DYE INTO BLADDER. UNFORTUNATELY, SUBSEQUENT ANTEGRADE SHOWED NO PASSAGE OF CONTRAST. URETEROSCOPY PERFORMED AT PRIOR SHOWED WIDE OPEN ANASTOMOSIS. HE HAD A RESONANCE STENT PLACED AND HAS TOLERATED IT WELL. WE AGREED TO PROCEED WITH CYSTOSCOPY, RIGHT RETROGRADE PYELOGRAM, RIGHT URETEROSCOPY, POSSIBLE LASER INCISION OF STRICTURE EXCHANGE OF RIGHT RESONANCE STENT, THE PROCEDURE WAS EXPLAINED IN DETAIL AND THE RISKS/BENEFITS/ALTERNATIVES WERE REVIEWED. I DISCUSSED IN FULL DETAIL ALL THE POSSIBLE COMPLICATIONS, RISK FACTORS, POSSIBLE NEED FOR ADDITIONAL PROCEDURES, RISK OF BLEEDING, RISKS THAT COME WITH ANESTHESIA (INCLUDING HEART ATTACK, STROKE, PULMONARY COMPLICATIONS, OR DEATH), ANTICIPATED RECOVERY TIME, AND I TOOK AMPLE TIME TO ENSURE ALL QUESTIONS WERE ANSWERED TO THE SATISFACTION OF THE PATIENT AND ALL PARTIES PRESENT DURING THIS DISCUSSION.¿

Description of Event or Problem · 0

SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 04-DEC-2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1504336 RESONANCE STENT SET FAD - STENT, URETERAL FAD COOK IRELAND LTD G34110 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male Required Intervention