FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-TSHR

MDR report key: 23349700 · Received October 21, 2025

Report

Report Number
1823260-2025-04275
Event Type
Malfunction
Date Received
October 21, 2025
Date of Event
September 29, 2025
Report Date
December 11, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JZO
UDI-DI
07613336160305
PMA / PMN Number
K193313
Removal / Correction Number
RES 98103
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ISSUES HAVE BEEN IDENTIFIED WITH THE ELECSYS ANTI-TSHR ASSAY, PARTICULARLY AROUND THE 1.75 IU/L MEDICAL DECISION POINT. ROCHE DIAGNOSTICS HAS ISSUED AN URGENT MEDICAL DEVICE CORRECTION FOR THIS ISSUE, ENTITLED "ELECSYS® ANTI-TSHR ASSAY ¿ DEVIATIONS ON COBAS® E 411, 402, AND 801 ANALYZERS." THIS CORRECTION HAS BEEN REPORTED TO FDA. ROCHE DIAGNOSTICS INVESTIGATED AND CONFIRMED THE CUSTOMERS' ALLEGATIONS. THE ROOT CAUSE FOR THE VARIATIONS IN THE ELECSYS ANTI-TSHR ASSAY WAS AN INSUFFICIENCY OF THE ASSAY'S ROUTINE PRODUCTION PROCESS. FOR THE PLATFORM-TO-PLATFORM DEVIATION (COBAS E 411 VS. COBAS E 402/E 801), DIFFERENCES AROSE DUE TO AN INSUFFICIENT ALIGNMENT CONCEPT BETWEEN THE PLATFORMS. THESE VARIATIONS WERE FURTHER INFLUENCED BY UPDATES TO THE ROUTINE PRODUCTION PROCESS FOR THIS ASSAY. FOR THE LOT-TO-LOT DEVIATION ON E 411, MEASURES TAKEN TO ENHANCE ROBUSTNESS LED TO ADDITIONAL SHIFTS. CUSTOMERS WERE ADVISED TO IMMEDIATELY DISCONTINUE USING AND DISCARD THE AFFECTED LOT NUMBERS: 840183 AND 874011 ON THE COBAS E402/E801 ANALYZER 840177 ON THE COBAS E411 ANALYZER CUSTOMERS WERE ADVISED TO SWITCH TO THE NEW UNAFFECTED LOT NUMBERS: 908953 ON THE COBAS E402/E801 ANALYZER 906872 ON THE COBAS E411 ANALYZER THE COBAS E601 AND E602 MODULES ARE NOT AFFECTED AND CAN BE USED WITH ALL CURRENTLY AVAILABLE LOTS AND WITH THE UPCOMING LOTS WITHOUT RESTRICTIONS.

Additional Manufacturer Narrative · 0

THE E411 ANALYZER SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

THE INITIAL REPORTER COMPLAINED OF HIGH RESULTS NOT CORRESPONDING TO THE CLINICAL PICTURE FOR MULTIPLE PATIENT SAMPLES TESTED FOR ELECSYS ANTI-TSHR (ANTI-TSHR) ON A COBAS E 411 ANALYZER (RACK SYSTEM) SINCE (B)(6) 2025. THE FOLLOWING EXAMPLES FOR 2 PATIENT SAMPLES TESTED ON (B)(6) 2025 WERE PROVIDED. PATIENT 1 RESULT FROM THE E411 ANALYZER WAS 1.75 U/L. THE RESULT AT ANOTHER LABORATORY USING AN UNSPECIFIED COBAS 8000 SYSTEM WAS 0.88 U/L. PATIENT 2 RESULT FROM THE E411 ANALYZER WAS 2.75 U/L. THE RESULT AT ANOTHER LABORATORY USING AN UNSPECIFIED COBAS 8000 SYSTEM WAS < 0.8 U/L. THE FIELD APPLICATION SPECIALIST (FAS) FOUND THAT QC WAS INITIALLY ACCEPTABLE BUT SHIFTED AFTER REPLACING THE REAGENT AND PERFORMING CALIBRATION. LIQUID FLOW CLEANING (LFC) AND MEASURING CELL PREPARATION WERE PERFORMED. CALIBRATION AND QC WERE COMPLETED, AND PATIENT SAMPLES WERE RUN. THE ISSUE WAS NOT RESOLVED. QC WAS WITHIN RANGE BUT WAS TRENDING HIGH. PATIENT RESULTS WERE ALSO HIGH. THE FIELD SERVICE ENGINEER (FSE) FOUND THAT QC FOR SOME TESTS WAS OUTSIDE OF THE ACCEPTABLE RANGE. THE MEASURING CELL WAS REPLACED. A CELL BLANK AND CALIBRATIONS WERE COMPLETED. AFTER THE MEASURING CELL WAS REPLACED, THE CUSTOMER COMPLAINED OF ANOTHER QUESTIONABLE RESULT. ON (B)(6) 2025 PATIENT 3 RESULT FROM THE E411 ANALYZER WAS 2.37 U/L. THE RESULT USING AN UNSPECIFIED ELECSYS REAGENT AT ANOTHER LABORATORY WAS 0.90 U/L. THE LOWER RESULTS WERE BELIEVED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
841582 ELECSYS ANTI-TSHR THYROID AUTOANTIBODY IMMUNOLOGICAL TEST SYSTEM JZO ROCHE DIAGNOSTICS 84017701 07613336160305

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown