UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-04716
- Event Type
- Malfunction
- Date Received
- October 21, 2011
- Date of Event
- April 16, 2010
- Report Date
- April 22, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SERVICE WAS NOT DISPATCHED AS THE CUSTOMER DID NOT QUESTION SYSTEM PERFORMANCE. THE ARCHIVE DATA FILES PROVIDED BY THE CUSTOMER INDICATED EVIDENCE OF PACK SHARING BETWEEN THE CUSTOMER'S TWO UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. BECKMAN COULTER CUSTOMER TECHNICAL SUPPORT (CTS) REVIEWED PACK LOADING AND SHARING PROTOCOLS WITH THE CUSTOMER. REAGENT PACK SHARING HAD OCCURRED AND PRODUCED ERRONEOUS RESULTS DUE TO OPERATOR ERROR. REAGENT PACK SHARING BETWEEN SYSTEMS CAN POTENTIALLY CAUSE INACCURATE RESULTS. REAGENT PACKS CANNOT BE SHARED BETWEEN SYSTEMS AS INDICATED IN THE LABELING. THE CUSTOMER ACKNOWLEDGED AND RE-EDUCATED THE OPERATORS THAT REAGENT PACK SHARING IS NOT PERMITTED ON THE ACCESS SYSTEMS. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS CONDUCTED FROM (B)(4) 2008 THROUGH (B)(4) 2010 FOR ADDITIONAL REPORTABLE EVENTS.
THE CUSTOMER REPORTED ERRONEOUSLY LOW THYROID-STIMULATING HORMONE (TSH), BELOW NORMAL CLINICAL RANGE, FOR TWENTY-THREE PTS' RESULTS, INVOLVING UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. SUBSEQUENT TESTING ON AN ALTERNATE UNICEL DXI 800 SYSTEM PRODUCED HIGHER RESULTS, IN BETTER CORRELATION WITH THE PTS' CLINICAL CONDITION. THE ERRONEOUS RESULTS WERE NOT RELEASED OUT OF THE LAB. THERE WAS NO PT INJURY OR CHANGE IN PT TREATMENT ASSOCIATED WITH THIS EVENT. BECKMAN COULTER, INC. REQUESTED FOR THE CUSTOMER'S ARCHIVE DATA FILES FOR FURTHER REVIEW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | ACCESS HYPERSENSITIVE HTSH |