FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,13

MDR report key: 23347867 · Received October 21, 2025

Report

Report Number
0001038806-2025-02574
Event Type
Injury
Date Received
October 21, 2025
Date of Event
September 17, 2025
Report Date
October 21, 2025
Manufacturer
ZIMVIE US CORP LLC
Product Code
DZE
UDI-DI
00889024019249
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER: (B)(4). A4: WEIGHT UNKNOWN / NOT PROVIDED. E1: FIRST/GIVEN NAME UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K072589. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #10 WAS REMOVED DUE TO BONE LOSS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
975984 IMP,TSV,4.1MM,SBM,13 DENTAL IMPLANT DZE ZIMVIE US CORP LLC 1293455 00889024019249

Patients

Seq Age Sex Outcome Treatment
1 45 YR Male Required Intervention