FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 23347359 · Received October 21, 2025

Report

Report Number
1723170-2025-03501
Event Type
Malfunction
Date Received
October 21, 2025
Date of Event
September 24, 2025
Report Date
October 21, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00643169722187
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821R, SERIAL/LOT #: (B)(6) UDI#: (B)(4) H3: THE CAMERA, LOT NUMBER: P902027, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED POSITIONING SENSOR UNIT (PSU) CONTAINED SCRATCHES ON THE HOUSING & LENSES. A CHECK OF THE EVENT LOG REVEALED INTERMITTENT FIRMWARE INCOMPATIBILITY AND INTERMITTENT ILLUMINATOR CURRENT LOW. THERE WAS ALSO A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED MESSAGES. THE PSU FAILED AN ACCURACY TEST (AAK) AT .656MM WITH A PASSING THRESHOLD OF .250MM. H6: CODES B01, C02, C08 AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED PRIOR TO A PROCEDURE. IT WAS REPORTED THAT WHILE SETTING UP FOR A CASE TODAY, THE SITE RECEIVED A LOCALIZER FAULTED MESSAGE IN THE APPLICATION. TROUBLESHOOTING WAS PERFORMED. THE NETWORK DEVICE INTERFACE (NDI) TOOLBOX REFLECTED A CMOS BATTERY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
879054 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. 9735665 00643169722187

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."