FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 23345966
·
Received October 21, 2025
Report
- Report Number
- 3008642652-2025-10126
- Event Type
- Malfunction
- Date Received
- October 21, 2025
- Date of Event
- October 17, 2025
- Report Date
- October 21, 2025
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005135
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
DEVICE EVALUATION OF ELECTRODE BELT HAS BEEN COMPLETED. THE REPORTED PROBLEM (ECGS NON-FUNCTIONAL) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT UNABLE TO COMPLETE AN ECG FALL-OFF TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A BROKEN BROWN (-5V) WIRE NEAR THE JST CONNECTOR IN THE TRUNK CABLE. THE ROOT CAUSE FOR THE BROKEN WIRE COULD NOT BE POSITIVELY IDENTIFIED. THERE WAS NO ADVERSE EVENT THAT RESULTED FROM THE DAMAGED BELT.
Description of Event or Problem · 0
A US DISTRIBUTOR REPORTED THAT AN ELECTRODE BELT HAD NON-FUNCTIONAL ECGS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2283644 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | LIFEVEST WCD 4000 SYSTEM | 00855778005135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |