FDA Adverse Event Malfunction Summary report: N

CD HORIZON SOLERA 5.5/6.0

MDR report key: 23345133 · Received October 21, 2025

Report

Report Number
3003120897-2025-00418
Event Type
Malfunction
Date Received
October 21, 2025
Date of Event
September 25, 2025
Report Date
October 21, 2025
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
HXX
UDI-DI
00613994429742
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS: PART # 5480004V; LOT # K24E3180. VISUAL EXAMINATION CONFIRMED THE T-25 DRIVER TIP HAS BROKEN. OPTICAL INSPECTION OF THE FRACTURE SURFACE REVEALED A FAIRLY FLAT DUCTILE FRACTURE WITH CIRCULAR MATERIAL FLOW. THE DAMAGE TO THE DRIVER IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING DEVICES USED IN POSTERIOR LUMB AR. IT WAS REPORTED THAT THE DRIVERS AND THE OBTURATOR WERE BROKEN. NO PATIENT WAS INVOLVED IN THIS EVENT. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
801314 CD HORIZON SOLERA 5.5/6.0 SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC. 5480004V K24E3180 00613994429742

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown