FDA Adverse Event Injury Summary report: N

SMR SYSTEM

MDR report key: 23345068 · Received October 21, 2025

Report

Report Number
3008021110-2025-00143
Event Type
Injury
Date Received
October 21, 2025
Date of Event
October 17, 2025
Report Date
October 20, 2025
Manufacturer
LIMACORPORATE S.P.A
Product Code
KWT
UDI-DI
08033390274296
PMA / PMN Number
K100858
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CHECKING THE DHRS OF THE INVOLVED LOT #1802753 (STER. #(B)(4)), NO PRE-EXISTING ANOMALY WAS FOUND ON (B)(4) DEVICES MANUFACTURED WITH THAT LOT AND STER. ACCORDING TO OUR RECORDS THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT # AND STER. #. THIS IS INITIAL MDR REPORT THE MANUFACTURER WILL SUBMIT A FINAL MDR REPORT WHEN THE INVESTIGATIONS ARE CLOSED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORDS AND STERILIZATION DOCUMENTATION WAS PERFORMED FOR THE INVOLVED LOT: · LOT #1802753 (STER. #(B)(4)): NO ANOMALIES, DEVIATIONS, OR NONCONFORMITIES WERE IDENTIFIED DURING MANUFACTURING OR STERILIZATION. A TOTAL OF (B)(4) DEVICES WERE MANUFACTURED AND STERILIZED UNDER THIS LOT. ACCORDING TO LIMACORPORATE RECORDS, THIS REPRESENTS THE FIRST AND ONLY COMPLAINT RECEIVED FOR THE INVOLVED LOT AND STERILIZATION NUMBER. DEVICE ANALYSIS THE EXPLANTED DEVICES WERE NOT RETURNED TO LIMACORPORATE FOR FURTHER PHYSICAL ANALYSIS. HOWEVER, PRE-REVISION X-RAYS AND PHOTOGRAPHS OF THE EXPLANTED DEVICES WERE PROVIDED BY THE COMPLAINT SOURCE. THIS DOCUMENTATION WAS REVIEWED INTERNALLY AND SUBSEQUENTLY SHARED WITH A MEDICAL EXPERT FOR A CLINICAL ASSESSMENT OF THE EVENT. THE EXPERT'S EVALUATION STATES: "THE CASE IS A ROTATOR CUFF FAILURE. THERE IS NO SIGN FOR IMPLANT RELATED FAILURE. IT IS A FATEFUL COURSE OF EVENTS." BASED ON THE AVAILABLE RADIOGRAPHIC DOCUMENTATION AND THE MEDICAL EXPERT'S ASSESSMENT, NO EVIDENCE SUGGESTIVE OF IMPLANT MALFUNCTION, MATERIAL FAILURE, OR DESIGN-RELATED ISSUES WAS IDENTIFIED. CONSIDERING THAT THE EXACT ROOT CAUSE OF THE EVENT CANNOT BE CONCLUSIVELY DETERMINED DUE TO THE LIMITED CLINICAL AND TECHNICAL INFORMATION AVAILABLE AND THE ABSENCE OF THE EXPLANTED DEVICES FOR PHYSICAL EXAMINATION, BASED ON THE FOLLOWING CONSIDERATIONS: · THE REVIEW OF THE DEVICE HISTORY RECORDS AND STERILIZATION DOCUMENTATION FOR THE INVOLVED LOT AND STERILIZATION NUMBER DID NOT IDENTIFY ANY ANOMALIES OR NONCONFORMITIES. · THIS IS THE FIRST AND ONLY COMPLAINT REPORTED FOR THE INVOLVED LOT AND STERILIZATION #; · THE MEDICAL EXPERT ASSESSMENT DID NOT IDENTIFY ANY IMPLANT-RELATED FAILURE AND ATTRIBUTED THE EVENT PRIMARILY TO ROTATOR CUFF FAILURE. · NO ADDITIONAL CLINICAL, RADIOGRAPHIC, OR TECHNICAL DATA ARE AVAILABLE TO SUPPORT FURTHER INVESTIGATION. THE EVENT IS ASSESSED AS NOT PRODUCT-RELATED. PMS DATA: 4 SIMILAR EVENTS FROM 9 UNITS SOLD IN AUSTRALIA (B)(4) IN THE PAST 3 YEARS. 8 SIMILAR EVENTS FROM (B)(4) UNITS SOLD WORLDWIDE (B)(4) IN THE PAST 3 YEARS. BASED ON THE ROOT CAUSE ANALYSIS PERFORMED AND ACCORDING TO THE RELEVANT PMS DATA, NO CORRECTIVE ACTIONS ARE REQUIRED FOR THIS SPECIFIC CASE. LIMACORPORATE CONTINUES MONITORING THE MARKET TO PROMPTLY DETECT ANY FURTHER SIMILAR ISSUE. THIS IS FINAL MDR REPORT.

Description of Event or Problem · 0

REVISION SURGERY PERFORMED ON (B)(6) 2025 DUE CUFF FAILURE. THE FOLLOWING SMR STEMLESS ANATOMIC IMPLANTS WERE REMOVED AND REVERSE FROM STRYKER WAS IMPLANTED: · SMR HUMERAL HEAD Ø52 MM (PRODUCT CODE 1322.09.520, LOT # 1802753 - STER. (B)(4)) SOLD IN US. · SMR STEMLESS - NEUTRAL ADAPTOR (PRODUCT CODE 1335.15.200, LOT # 1713952 - STER. (B)(4)) SOLD IN US. · SMR STEMLESS - STEML. CORE #M (PRODUCT CODE 1355.14.235, LOT # 1714291 - STER. (B)(4)) SOLD IN US · CEMENTED GLENOID 3 PEGS SMALL (PRODUCT CODE 1379.51.020, LOT # 15AT1PG - STER. (B)(4)) PREVIOUS SURGERY WAS PERFROMED ON (B)(6) 2018. PATIENT IS A MALE, 67 YEARS OLD. EVENT HAPPENED IN AUSTRALIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2355459 SMR SYSTEM SMR HUMERAL HEAD Ø52 MM KWT LIMACORPORATE S.P.A 1802753 08033390274296

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other