FDA Adverse Event Injury Summary report: N

KIT MMT-1761KES 670G V4.11 JBL MG

MDR report key: 23345038 · Received October 21, 2025

Report

Report Number
2032227-2025-287300
Event Type
Injury
Date Received
October 21, 2025
Date of Event
October 2, 2025
Report Date
October 20, 2025
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000367657
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UY
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER ALSO REPORTED AN INSULIN FLOW BLOCKED ALARM. BLOOD GLUCOSE VALUE AT THE TIME OF EVENT WAS 400 MG/DL. THE CUSTOMER WAS TREATED WITH A MANUAL INJECTION. THE EVENT INVOLVED PRODUCT(S) MMT-1761KES. TROUBLESHOOTING WAS PERFORMED FOR HIGH BLOOD GLUCOSE EVENT AND THE CUSTOMER WAS ADVISED TO CONSIDER CHANGING THE INSULIN, RESERVOIR AND INFUSION SET. THE CUSTOMER WAS USING THE AUTO MODE OR SMART GUARD FEATURE AT THE TIME OF THE EVENT. THE CUSTOMER WAS USING THE INSULIN PUMP SYSTEM WITHIN 48 HOURS OF THE REPORTED EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-1761KES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2285414 KIT MMT-1761KES 670G V4.11 JBL MG AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1761KES HG5AYFV 000000763000367657

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other FRN-MMT-332A-RSVR, UNOMED SET, OZP-MMT-7020- SNSR