FDA Adverse Event Malfunction Summary report: N

SECHRIST INDUSTRIES INC.

MDR report key: 23342249 · Received October 20, 2025

Report

Report Number
2020676-2025-00025
Event Type
Malfunction
Date Received
October 20, 2025
Date of Event
September 4, 2025
Report Date
October 20, 2025
Manufacturer
SECHRIST INDUSTRIES INC.
Product Code
DTX
UDI-DI
00899660002345
PMA / PMN Number
K023745
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A GETINGE FIELD SERVICE TECHNICIAN WAS SENT FOR INVESTIGATION AND REPAIR ON 2025-09-29. THE SECHRIST GAS MIXER WAS REPLACED. THE AIR-OXYGEN MIXER MUST BE REPLACED WITH A NEW ONE. THE AIR-OXYGEN MIXER IS UNSUITABLE FOR CLINICAL USE. THE REPORTED FAILURE DID NOT CAUSE OR CONTRIBUTE TO SERIOUS INJURY OR DEATH. SINCE NO SYSTEMIC ISSUE WAS DETERMINED NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. AT THIS TIME THERE IS NO EVIDENCE THAT A MANUFACTURING NON-CONFORMITY CONTRIBUTED TO THE REPORTED COMPLAINT, AND THE INSTRUCTIONS FOR USE WERE REVIEWED AND DETERMINED TO PROVIDE ADEQUATE INSTRUCTIONS AND WARNINGS FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE. THE IFU STATES, "TO ENSURE PROPER FUNCTION AND ACCURACY, THE SECHRIST AIR/OXYGEN GAS MIXERS MUST BE THOROUGHLY OVERHAULED EVERY TWO (2) YEARS. TO MAINTAIN THE PRODUCT WARRANTY, THE OVERHAUL MUST BE PERFORMED BY SECHRIST INDUSTRIES OR BY SECHRIST AUTHORIZED PERSONNEL." THEREFORE, NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE NECESSARY. ALL COMPLAINTS ARE TRENDED AND REVIEWED BY MANAGEMENT ON A MONTHLY BASIS. AS PART OF THIS MONTHLY REVIEW, ANY EXCURSION ABOVE THE CONTROL LIMITS FOR THIS FAILURE MODE WILL BE ASSESSED, DOCUMENTED AND ACTED UPON AS WARRANTED. MANUFACTURER REFERENCE FILE NO. (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT BLOOD HAS ENTERED THE GAS MIXER. NO HARM TO ANY PERSON WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2452412 SECHRIST INDUSTRIES INC. GAS CONTROL UNIT, CARDIOPULMONARY BYPASS DTX SECHRIST INDUSTRIES INC. 3500CP-G 00899660002345

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown