FDA Adverse Event Malfunction Summary report: N

ARCHITECT FREE T4

MDR report key: 23342243 · Received October 20, 2025

Report

Report Number
3005094123-2025-00521
Event Type
Malfunction
Date Received
October 20, 2025
Date of Event
October 10, 2025
Report Date
December 22, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
CEC
UDI-DI
00380740173678
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A COMPLAINT INVESTIGATION WAS CONDUCTED FOR THE ARCHITECT FREE T4 REAGENT LOT 73030UD02. A REVIEW OF TRACKING AND TRENDING DID IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR THE ARCHITECT FREE T4 ASSAY, HOWEVER, NO TRENDS WERE IDENTIFIED FOR THE COMPLAINT LOT. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT LOT NUMBER AND ISSUE. ACCURACY TESTING WAS COMPLETED WITH A RETAINED KIT OF THE COMPLAINT LOT. ALL VALIDITY AND ACCEPTANCE CRITERIA WERE MET DURING COMPLETION OF THE PROTOCOL, DEMONSTRATING THAT THE LOT IS PERFORMING AS EXPECTED. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT FREE T4 REAGENT LOT 73030UD02.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K65-77 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 07K65, 510K K173122.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON AN ARCHITECT I2000SR ANALYZER FOR A 38-YEAR-OLD MALE PATIENT. SID (B)(6) INITIAL RESULT = 26.66 PMOL/L, REPEAT RESULTS = 13.39 PMOL/L (REFERENCE RANGE 9.09-19.04 PMOL/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT FREE T4 RESULT GENERATED ON AN ARCHITECT I2000SR ANALYZER FOR A 38-YEAR-OLD MALE PATIENT. SID (B)(6) INITIAL RESULT = 26.66 PMOL/L, REPEAT RESULTS = 13.39 PMOL/L (REFERENCE RANGE 9.09-19.04 PMOL/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2452406 ARCHITECT FREE T4 RADIOIMMUNOASSAY, FREE THYROXINE CEC ABBOTT IRELAND DIAGNOSTICS DIVISION 73030UD02 00380740173678

Patients

Seq Age Sex Outcome Treatment
1 38 YR Male ARC I2000SR INST, 03M74-02, (B)(6).