COULTER® ACT DIFF ANALYZER
Report
- Report Number
- 1061932-2011-02159
- Event Type
- Malfunction
- Date Received
- November 12, 2011
- Date of Event
- October 14, 2011
- Report Date
- October 14, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- K973634
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(4) 2011, BEC CTS (CUSTOMER TECHNICAL SUPPORT) HAD THE CUSTOMER INSPECT THE RINSE BLOCK AND TUBING. THE CUSTOMER FOUND AND REMOVED A CLOT AND THEN RAN STARTUP AND SAMPLE TESTED COMPARED TO PREVIOUS RUN. SERVICE WAS NOT DISPATCHED. WHEN THE CUSTOMER CALLED BACK ON (B)(6) 2011 WITH REGARD TO THE WBC BATH NOT DRAINING USING DILUTER FUNCTIONS, A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON-SITE AND FOUND A CLOT IN THE SYSTEM UPON WHICH FSE REPLACED THE WBC BATH AND DRAIN SOLENOID VALVE. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES. THE ROOT CAUSE FOR THIS EVENT IS ATTRIBUTED TO THE CLOT. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT THE WBC BATH IN THE COULTER ACT DIFF ANALYZER WAS NOT DRAINING AND WAS OVERFLOWING ONTO THE COUNTER. THE OPERATOR WAS WEARING GLOVES AND A LAB COAT AT THE TIME OF THE EVENT. NO INJURY OR EXPOSURE WAS REPORTED AND MEDICAL ATTENTION WAS NOT SOUGHT. THE CUSTOMER HAD CALLED BEC A DAY BEFORE THIS LEAK ((B)(6) 2011) STATING THAT THE INSTRUMENT WAS DRIPPING FROM THE PROBE (THIS EVENT HAS BEEN DOCUMENTED IN A SEPARATE MDR REPORT #1061932-2011-02158).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® ACT DIFF ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER INC. | ACT DIFF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |