MDR report key: 2333760 · Received November 11, 2011

Report

Report Number
6000001-2011-37259
Event Type
Malfunction
Date Received
November 11, 2011
Date of Event
October 21, 2011
Report Date
October 21, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE OR THE DEVICE IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 589 WAS CONFIRMED DURING SERVICE EVALUATION. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES AS A RESULT OF USE ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. THE USER INTERFACE MODULE SOFTWARE VERSION IS 6.13.92. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT A COLLEAGUE INFUSION PUMP FOR FAILURE CODE 589:317:1061. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. PATIENT INVOLVEMENT IS UNKNOWN. IT IS UNKNOWN IF THERE WAS AN ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1