Narratives
(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE OR THE DEVICE IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 589 WAS CONFIRMED DURING SERVICE EVALUATION. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES AS A RESULT OF USE ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THE REPORTED CONDITION. THE USER INTERFACE MODULE SOFTWARE VERSION IS 6.13.92. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT A COLLEAGUE INFUSION PUMP FOR FAILURE CODE 589:317:1061. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. PATIENT INVOLVEMENT IS UNKNOWN. IT IS UNKNOWN IF THERE WAS AN ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.