FDA Adverse Event Malfunction Summary report: N

HD CAMERA HEAD

MDR report key: 23336046 · Received October 20, 2025

Report

Report Number
3002808148-2025-18499
Event Type
Malfunction
Date Received
October 20, 2025
Date of Event
September 17, 2025
Report Date
December 29, 2025
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
PMA / PMN Number
K102059
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, H2, H3, H6 AND H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: CABLE WATER INGRESS AND IMAGING WATER INGRESS. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THIS COMPLAINT IS EXPECTED OR RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT IN THE SUBJECT DEVICE THE EYEPIECE LENS COMES OFF. THE ISSUE OCCURRED DURING LAB DEMONSTRATION. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340339 HD CAMERA HEAD AUTOCLAVABLE CAMERA HEAD FET SHIRAKAWA OLYMPUS CO., LTD. CH-S190-08-LB

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown