FDA Adverse Event Injury Summary report: N

SMARTDRIVE

MDR report key: 23335210 · Received October 20, 2025

Report

Report Number
3008370857-2025-00208
Event Type
Injury
Date Received
October 20, 2025
Date of Event
September 4, 2025
Report Date
October 20, 2025
Manufacturer
MAX MOBILITY LLC
Product Code
ITI
UDI-DI
00861896000310
PMA / PMN Number
K151199
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PERMOBIL AUSTRALIA RECEIVED A REPORT CLAIMING THE SPEED CONTROL DIAL HAVING INITIATED A DRIVE COMMAND WITHOUT ANY PHYSICAL MANIPULATION FROM THE END-USER. THIS ACTION REPORTEDLY FORCED THE END-USER TO MANEUVER THEIR WHEELCHAIR INTO AN IMMOVABLE OBJECT WHICH RESULTED IN THE END-USER LOSING UPRIGHT POSITIONING AND FALLING OUT OF THE SEATING WHERE THEY SUSTAINED BI-LATERAL FRACTURS TO THEIR LEGS. DUE TO AN INCREASE IN REPORTS OF SIMILAR FAILURES, CAPA 25-007 WAS IMPLEMENTED. INVESTIGATION FOUND A POTENTIAL ISSUE WHERE IF AN INTERMITTENT LOSS OF CONNECTION OCCURS BETWEEN THE SPEED CONTROL DIAL AND SMARTDRIVE ELECTRONICS, AN INVOLUNTARY ACTIVATION OF THE DRIVE MOTOR COULD OCCUR. FURTHER ENGINEERING ANALYSIS CONCLUDED THAT THIS ANOMALY WOULD ONLY OCCUR IF THE SPEED CONTROL DIAL WAS POWERED ON AND IN A FORWARD ROTATED DRIVE POSITION BUT IN STAND-BY MODE. IF THE CONNECTION IS LOST AND RE-INTRODUCED WHILE IN THIS POSITION, THE ELECTRONICS COULD POTENTIALLY INTERPRET THE RE-INTRODUCTION OF SIGNAL AS BEING A NEW COMMAND TO ENGAGE THE DRIVE MOTOR. AS THIS TYPE OF FAILURE CAN BE REPRODUCED AND HAS THE POTENTIAL TO OCCUR WITH ALL VERSIONS OF SPEED CONTROL DIAL, AS PART OF THE CAPA, PERMOBIL HAS IMPLEMENTED A PRODUCT RECALL FOR ALL SPEED CONTROL DIALS IN THE MARKET, Z-2538-2025, Z-2539-2025, Z-2540-2025. THE END-USER WILL BE PROVIDED WITH A SWITCH CONTROL BUTTON TO REPLACE THE AFFECTED SPEED CONTROL DIAL.

Description of Event or Problem · 0

REPORT CLAIMS THE SPEED CONTROL DIAL ENGAGED A DRIVE COMMAND ON THE SMARTDRIVE UNIT WITHOUT ANY PHYSICAL MANIPULATION FROM THE END-USER. THIS ACTION REPORTEDLY CAUSED THE USER TO COLLIDE INTO AN IMMOVEABLE OBJECT WHICH RESULTED IN A SERIOUS INJURY REQUIRING MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2820795 SMARTDRIVE SMARTDRIVE ITI MAX MOBILITY LLC MX2+ 00861896000310

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization