FDA Adverse Event Malfunction Summary report: N

VACUETTE® TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3.2% 13X75 HIGH ALTITUDE

MDR report key: 23328300 · Received October 17, 2025

Report

Report Number
1125230-2025-00048
Event Type
Malfunction
Date Received
October 17, 2025
Report Date
October 17, 2025
Manufacturer
GREINER BIO-ONE NA INC.
Product Code
GIM
PMA / PMN Number
K971221
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT (B)(4): RECEIVED 1 RACK 454351/B2504346 FOR EVALUATION. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED WITH NO DEVIATIONS OBSERVED. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE COMPLAINT COULD NOT BE DUPLICATED.

Additional Manufacturer Narrative · 0

COMPLAINT (B)(4). CUSTOME SAMPLES HAVE BEEN REQUESTED BUT NOT YET RECEIVED AT THE TIME OF THIS REPORT. IF AND WHEN SAMPLES ARRIVE, SAMPLES WILL BE EVALUATED AS PART OF THE COMPLAINT INVESTIGATION. UPON COMPLETION OF THE COMPLAINT INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

CUSTOMER STATES TUBES ARE UNDERFILLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267679 VACUETTE® TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3.2% 13X75 HIGH ALTITUDE EVACUATED BLOOD COLLECTION TUBE GIM GREINER BIO-ONE NA INC. 454351 B2504346

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown