VACUETTE® TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3.2% 13X75 HIGH ALTITUDE
Report
- Report Number
- 1125230-2025-00048
- Event Type
- Malfunction
- Date Received
- October 17, 2025
- Report Date
- October 17, 2025
- Manufacturer
- GREINER BIO-ONE NA INC.
- Product Code
- GIM
- PMA / PMN Number
- K971221
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
COMPLAINT (B)(4): RECEIVED 1 RACK 454351/B2504346 FOR EVALUATION. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED WITH NO DEVIATIONS OBSERVED. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE COMPLAINT COULD NOT BE DUPLICATED.
COMPLAINT (B)(4). CUSTOME SAMPLES HAVE BEEN REQUESTED BUT NOT YET RECEIVED AT THE TIME OF THIS REPORT. IF AND WHEN SAMPLES ARRIVE, SAMPLES WILL BE EVALUATED AS PART OF THE COMPLAINT INVESTIGATION. UPON COMPLETION OF THE COMPLAINT INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
CUSTOMER STATES TUBES ARE UNDERFILLING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 267679 | VACUETTE® TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3.2% 13X75 HIGH ALTITUDE | EVACUATED BLOOD COLLECTION TUBE | GIM | GREINER BIO-ONE NA INC. | 454351 | B2504346 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |