NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2011-00699
- Event Type
- Malfunction
- Date Received
- November 10, 2011
- Report Date
- October 13, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT DEVICE WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION, ALTHOUGH THE CUSTOMER WAS ASKED TO RETURN THE FAULTY PART. RESULTS: THE CUSTOMER HAS INFORMED US THAT THE THEY HAVE REPLACED THE FAULTY MANOMETER WITH A WORKING SPARE PART AND HAVE DISCARDED THE FAULTY PART. WHEN ASKED THEY DID NOT SUPPLY US WITH ANY INFORMATION AS TO THE REPORTED FAULT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090610. CONCLUSION: WHILE WE WERE UNABLE TO ASCERTAIN THE EXACT NATURE OF THE MANOMETER ISSUE, THE MOST COMMON CAUSE OF PROBLEMS WITH THE MANOMETER IS DAMAGE DUE TO THE NEOPUFF RECEIVING A SUBSTANTIAL IMPACT. THE NEOPUFF TECHNICAL MANUAL STATES THAT "DROPPING OF THE F&P NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." THE USER INSTRUCTIONS REQUIRE THAT THE NEOPUFF BE TESTED BY QUALIFIED PERSONNEL OR AN AUTHORISED FISHER & PAYKEL HEALTHCARE REPRESENTATIVE PRIOR TO EACH USE TO ENSURE FUNCTIONALITY. THE OPERATING MANUAL INSTRUCTS THE USER TO CARRY OUT A SET-UP PROCEDURE "PRIOR TO EVERY USE OF THE NEOPUFF TO ENSURE THAT THE DEVICE IS FUNCTIONING CORRECTLY." THE SET-UP PROCEDURE STATES THE USER MUST "CHECK MANOMETER READS ZERO WITH NO GAS FLOW. IF NOT, THE MANOMETER REQUIRES CALIBRATION" PER THE TECHNICAL MANUAL. IN ADDITION, THE NEOPUFF SET-UP PROCEDURE STATES THE USER MUST CHECK THE SETTINGS BY TESTING THE PRESSURE OUTPUT FROM THE NEOPUFF AT THE DESIRED FLOW RATE "PRIOR TO EVERY USE OF THE NEOPUFF TO ENSURE THAT THE DEVICE IS FUNCTIONING CORRECTLY."
A DISTRIBUTOR IN (B)(6) REPORTED THAT THE MANOMETER OF AN RD900 INFANT RESUSCITATOR NEEDED REPAIR. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 090610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |