KLIMT EXPANDABLE LIF CAGE
Report
- Report Number
- 3009051471-2025-00020
- Event Type
- Malfunction
- Date Received
- October 17, 2025
- Date of Event
- September 23, 2025
- Report Date
- October 17, 2025
- Manufacturer
- CTL MEDICAL CORPORATION
- Product Code
- MAX
- PMA / PMN Number
- K233947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
NO PATIENT INJURY OR SURGICAL DELAY WAS REPORTED. DURING SURGERY, THE TORQUE-LIMITING HANDLE SUCCESSFULLY REACHED ITS DESIGNED EXPANSION CAPACITY. THE SURGEON, HOWEVER, CONTINUED TO ROTATE THE HANDLE MULTIPLE TIMES AFTER THE TORQUE LIMITS. THIS REPETITION EXCEEDED THE MECHANICAL DESIGN THRESHOLD OF THE DRIVER TIP, RESULTING IN TORSIONAL OVERLOAD AND HEARING OF THE TIP. THE OBSERVED FAILURE IS CONSISTENT WITH MISUSE AND OVER-APPLICATION OF TORQUE BEYOND THE VALIDATED DESIGN PARAMETERS.
DURING A TLIF PROCEDURE: SURGEON COMPLETED ALL STEPS OF THE DISCECTOMY AND FIRST CAGE INSERTION (NON-CTL) AS NORMAL. WHEN HE WAS EXPANDING THE CAGE THE FIRST TIME, THE INSERTER DID NOT CATCH THE CONNECTION OF THE IMPLANT. ONCE WE RESOLVED THIS ISSUE, HE WAS ABLE TO EXPAND THE IMPLANT. UPON REACHING THE TORQUE LIMIT, THE REP INFORMED THE SURGEON THAT HE WAS FULLY EXPANDED TO THE CAPACITY OF THE TORQUE DRIVER. HE CONTINUED TO TURN THE TORQUE HANDLE AND IT DID SOUND LIKE THERE WAS SOME SLIGHT ADDITIONAL EXPANSION OCCURING WITH EACH TURN OF THE HANDLE BUT THE HANDLE TORQUED OUT SOON AFTER THAT. HE REPEATED THIS ABOUT 10-15 TIMES. THEN THE HANDLE TORQUED OUT AND STARTED SPINNING FREELY, INDICATING THE TIP HAD BEEN SHEARED OFF. HE PACKED THE SPACE WITH BONE GRAFT AFTER DETERMINING THE PIECE OF THE INSERTER COULD NOT BE REMOVED FROM THE INNER CONNECTION OF THE IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298664 | KLIMT EXPANDABLE LIF CAGE | EXPANSION DRIVER | MAX | CTL MEDICAL CORPORATION | 367.7508 | FG8AA* |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |