FDA Adverse Event Malfunction Summary report: N

TRILOGY 100 VENTILATOR

MDR report key: 23319389 · Received October 16, 2025

Report

Report Number
2518422-2025-049790
Event Type
Malfunction
Date Received
October 16, 2025
Date of Event
September 24, 2025
Report Date
October 17, 2025
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
UDI-DI
00606959015364
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 0

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP DURING TESTING FOR LOW OXYGEN FLOW. THE DEVICE'S INLET AIR PATH, AIR PATH FOAM, AND FLOW PATH WERE REPLACED TO ADDRESS THE ISSUE. THE DEVICE PASSED ALL TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2098216 TRILOGY 100 VENTILATOR VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054260 00606959015364

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown