PALODENT V3 UNIV 2 RING REFIL
Report
- Report Number
- 2515379-2025-00023
- Event Type
- Malfunction
- Date Received
- October 16, 2025
- Report Date
- October 21, 2025
- Manufacturer
- DENTSPLY LLC
- Product Code
- DZN
- UDI-DI
- D002659760V1
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, IF THIS MALFUNCTION RECURRED, IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803.
RETURN: 10-17-2025: PRODUCT NOT RETURNED, IMAGE IN CASE DEPICTS 1 V3 PALODENT UNIVERSAL BLUE RING (NEW AND IMPROVED V5 DESIGN) WITH BROKEN TYNE. OVER MOLDING DATE CODES VERIFIED ¿L¿ FOR DECEMBER AND ¿P¿ FOR 2024. DHR AND RETAIN EVALUATION WILL BE CONDUCTED. (NWV). RETAIN: 10-17-2025: FINAL PRODUCT RETAINS ARE NOT KEPT AS PER NORMAL PROCEDURE. RETAINS FROM OVER MOLDING ITEM#: 759870, LOT#¿S: 09261537, 09261538 & 09329082 WERE REVIEWED AND INSPECTED PER 0290-IP-7.5-60-58 AND WERE FOUND ACCEPTABLE. (NWV). DHR: 10-17-2025: DHR FOR ITEM#: 659760V, LOT#: 09323412 HAS BEEN PULLED, REVIEWED, AND ATTACHED TO THIS CASE. DHR REVIEW DID NOT INDICATE ANY PRODUCTION ISSUES WHILE PACKAGING/LABELING THE PALODENT V3 UNIV 2 RING REFILL. WORK ORDER: (B)(4) IS THE PACKAGING WORK ORDER WHICH UTILIZED 3 OVER-MOLDING OF THE SPRINGS TO RINGS PRODUCTION WORK ORDERS/RUNS OF ITEM: 759870 (V5 RING UNIVERSAL - PALODENT) LOTS IN WHICH WERE 09261537 (PRODUCED 12-2024), 09261538 (PRODUCED 12-2024) & 09329082 (PRODUCED 12-2024). DHRS FOR EACH MOLDING WORK ORDER HAVE ALSO BEEN PULLED, REVIEWED, AND ATTACHED TO THIS CASE. DHR REVIEW DID NOT IDENTIFY ANY ISSUED DURING PRODUCTION WITH ALL INSPECTIONS PERFORMED AND DEEMED ACCEPTABLE BY THE OPERATOR(S) AND QUALITY AS PER 0290-WI-7.5-60-14 & 0290-IP-7.5-60-58. (NWV).
IN THIS EVENT IT IS REPORTED THAT A PALODENT V3 UNIV 2 RING REFIL RING BROKE DURING USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2253304 | PALODENT V3 UNIV 2 RING REFIL | INSTRUMENTS, DENTAL HAND | DZN | DENTSPLY LLC | 9323412 | D002659760V1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |