FDA Adverse Event
Malfunction
Summary report: N
QUANTUM INTERSEGMENTAL TABLE
MDR report key: 23318345
·
Received October 16, 2025
Report
- Report Number
- 3007278668-2025-00003
- Event Type
- Malfunction
- Date Received
- October 16, 2025
- Date of Event
- September 22, 2025
- Report Date
- October 16, 2025
- Manufacturer
- PIVOTAL HEALTH SOLUTIONS, INC.
- Product Code
- ISA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
EVALUATION OF PHOTOS PROVIDED BY CUSTOMER OF PRIOR FAILURE DETERMINED IT WAS LIKELY THERE WAS MATERIAL BREAKDOWN OF HEAT COMPONENT CAUSED BY ACCELERATED WEAR AND REPEATED PINPOINT PRESSURE FROM PATIENT LOADING/UNLOADING RESULTING IN AN ELECTRICAL SHORT WITHIN THE INTERNAL COILS OF THE HEATING ELEMENT.
Description of Event or Problem · 0
A HEAT FUNCTION ON A QUANTUM TABLE MANUFACTURED IN 2010 MALFUNCTIONED AND OVERHEATED CAUSING THE VINYL TO GET HOT AND THE PATIENT FELT THE HEAT AND HAD A RED MARK ON THEIR ARM. THERE WAS NO PAIN, BURNING OR DISCOMFORT AND THE PATIENT DID NOT REQUIRE ANY FURTHER MEDICAL TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2254108 | QUANTUM INTERSEGMENTAL TABLE | QUANTUM | ISA | PIVOTAL HEALTH SOLUTIONS, INC. | Q400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |