FDA Adverse Event Malfunction Summary report: N

QUANTUM INTERSEGMENTAL TABLE

MDR report key: 23318345 · Received October 16, 2025

Report

Report Number
3007278668-2025-00003
Event Type
Malfunction
Date Received
October 16, 2025
Date of Event
September 22, 2025
Report Date
October 16, 2025
Manufacturer
PIVOTAL HEALTH SOLUTIONS, INC.
Product Code
ISA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EVALUATION OF PHOTOS PROVIDED BY CUSTOMER OF PRIOR FAILURE DETERMINED IT WAS LIKELY THERE WAS MATERIAL BREAKDOWN OF HEAT COMPONENT CAUSED BY ACCELERATED WEAR AND REPEATED PINPOINT PRESSURE FROM PATIENT LOADING/UNLOADING RESULTING IN AN ELECTRICAL SHORT WITHIN THE INTERNAL COILS OF THE HEATING ELEMENT.

Description of Event or Problem · 0

A HEAT FUNCTION ON A QUANTUM TABLE MANUFACTURED IN 2010 MALFUNCTIONED AND OVERHEATED CAUSING THE VINYL TO GET HOT AND THE PATIENT FELT THE HEAT AND HAD A RED MARK ON THEIR ARM. THERE WAS NO PAIN, BURNING OR DISCOMFORT AND THE PATIENT DID NOT REQUIRE ANY FURTHER MEDICAL TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2254108 QUANTUM INTERSEGMENTAL TABLE QUANTUM ISA PIVOTAL HEALTH SOLUTIONS, INC. Q400

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown