FISHER & PAYKEL HEALTHCARE
Report
- Report Number
- 9611451-2025-01069
- Event Type
- Malfunction
- Date Received
- October 16, 2025
- Report Date
- October 17, 2025
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(6). SECTION D4: UDI DETAILS ARE NOT AVAILABLE BASED ON THE TIME OF MANUFACTURING. SECTION G4: THE IW990AGU MANUAL-CONTROL WALL MOUNT INFANT WARMER IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K971441. METHOD: THE COMPLAINT IW990AGU WALL MOUNT INFANT WARMER WAS RECEIVED AT OUR FISHER & PAYKEL HEALTHCARE (F&P) SERVICE CENTRE IN GERMANY WHERE IT WAS VISUALLY INSPECTED AND SERVICED BY A TRAINED F&P SERVICE TECHNICIAN. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION PROVIDED BY THE F&P SERVICE TECHNICIAN, INFORMATION PROVIDED BY THE HEALTHCARE FACILITY AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: VISUAL INSPECTION OF THE SUBJECT DEVICE CONFIRMED THAT THE WARMER HEAD WAS DAMAGED. CONCLUSION: WE ARE UNABLE TO DETERMINE THE EXACT CAUSE OF THE REPORTED EVENT. THE USER MANUAL FOR THE IW990 INFANT WARMER STATES THE FOLLOWING: "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYDROCHLORIDES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE." "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES." THE DEVICE TECHNICAL/SERVICE MANUAL ALSO CONTAINS A CHECKLIST WHICH SPECIFIES THAT USERS PERFORM SAFETY, PERFORMANCE, AND FUNCTIONAL CHECKS AT LEAST ONCE A YEAR.
A HEALTHCARE FACILITY IN GERMANY REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE WARMER HEAD OF AN IW990AGU WALL MOUNT INFANT WARMER HAS CRACKS. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2694854 | FISHER & PAYKEL HEALTHCARE | INFANT WARMER | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW990AGU | 071119 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |