FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 23311274 · Received October 16, 2025

Report

Report Number
2647876-2025-00653
Event Type
Malfunction
Date Received
October 16, 2025
Date of Event
September 28, 2025
Report Date
November 17, 2025
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CATALOG 442021. BATCH NO. 5093639. CUSTOMER REPORTED A CONTAMINATION ISSUE WHILE USING BACTEC PRODUCT. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. PRODUCT INSERTS STATES THAT CARE MUST BE TAKEN TO PREVENT CONTAMINATION OF THE SAMPLE DURING COLLECTION AND INOCULATION INTO THE BD BACTEC¿ VIAL. PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSED SEPTUM, OR LEAKAGE. DO NOT USE ANY VIAL SHOWING EVIDENCE OF CONTAMINATION. A CONTAMINATED VIAL COULD CONTAIN POSITIVE PRESSURE. IF A CONTAMINATED VIAL IS USED FOR DIRECT DRAW, GAS OR CONTAMINATED CULTURE MEDIA COULD BE REFLUXED INTO THE PATIENT¿S VEIN. VIAL CONTAMINATION MAY NOT BE READILY APPARENT. PRIOR TO USE, THE USER SHOULD EXAMINE THE VIALS FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS THAT ARE CRACKED OR LEAKING, OR DISPLAY TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. THE SPECIMEN MUST BE COLLECTED USING STERILE TECHNIQUES TO REDUCE THE CHANCE OF CONTAMINATION. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 6: IT WAS REPORTED WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) THERE WAS BIOLOGICAL CONTAMINATION SEEN IN ONE SAMPLE IDENTIFIED AS BC BY ENVIRONMENTAL BACTERIA - PENICILLIUMS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

REPORT 6: IT WAS REPORTED WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) THERE WAS BIOLOGICAL CONTAMINATION SEEN IN ONE SAMPLE IDENTIFIED AS BC BY ENVIRONMENTAL BACTERIA - PENICILLIUMS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2664504 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 5093639 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown