FDA Adverse Event Injury Summary report: N

GLIDEWELL HT IMPLANT Ø5.0 X 10 MM

MDR report key: 23311033 · Received October 16, 2025

Report

Report Number
3011649314-2025-01218
Event Type
Injury
Date Received
October 16, 2025
Date of Event
July 14, 2025
Report Date
January 8, 2026
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN COMPLETED, AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR GLIDEWELL HT IMPLANT LOT# 6240877, AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF THE STOCK PRODUCT WAS PERFORMED FOR GLIDEWELL HT IMPLANT LOT# 6240877 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED, BUT NOT IN THE ORIGINAL PACKAGE. AN ABUTMENT WAS ATTACHED TO THE IMPLANT. THE IMPLANT WAS VERIFIED TO BE A GLIDEWELL HT IMPLANT Ø5.0 X 10 MM (70-1189-IMP0015) USING THE RADIOGRAPHIC TEMPLATE (MKT-013579 REV 1 PK-4500230-112023). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED, AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART, BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE: "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS TYPE III BONE QUALITY AND FAIR ORAL HYGIENE. IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "GLIDEWELL HT IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE" IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. THE RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATION, AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." THE MANUFACTURER'S INTERNAL REFERENCE NUMBER IS: (B)(4). ADDITIONAL INFORMATION: G3: "DATE RECEIVED BY MANUFACTURER" H3: "DEVICE EVALUATED BY MANUFACTURER" CORRECTIONS: B5: "DESCRIBE EVENT OR PROBLEM" B7: "OTHER RELEVANT HISTORY" H6: UPDATED "TYPE OF INVESTIGATION" CODE. H6: UPDATED "INVESTIGATION FINDINGS" CODE. H6: UPDATED "INVESTIGATION CONCLUSIONS" CODE.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. HOWEVER, THE EVALUATION OF THE DEVICE IS PENDING. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A GLIDEWELL HT IMPLANT WAS PLACED IN TOOTH POSITION #15 ON (B)(6) 2024 IN A 60-YEAR-OLD MALE PATIENT WITH TYPE 2 BONE QUALITY AND FAIR ORAL HYGIENE. THE IMPLANT WAS PLACED FOLLOWING IMMEDIATE EXTRACTION AND WAS NOT IMMEDIATELY LOADED. THE PATIENT'S OPPOSING ARCH CONSISTED OF NATURAL TEETH. ON (B)(6) 2025, PRIOR TO SECOND STAGE SURGERY, THE PATIENT PRESENTED WITH INFLAMMATION/INFECTION. THE IMPLANT FAILED TO OSSEOINTEGRATE AND WAS REMOVED ON (B)(6) 2025. PATIENT SYMPTOMS RESOLVED FOLLOWING IMPLANT REMOVAL. NO PERMANENT INJURY OCCURRED, AND NO ADDITIONAL MEDICAL OR SURGICAL PROCEDURES WERE REQUIRED. THE IMPLANT WAS REPLACED WITH A SIMILAR PRODUCT. NO ABNORMALITIES WERE NOTED WITH THE IMPLANT OR PACKAGING UPON VISUAL INSPECTION.

Description of Event or Problem · 0

THE PATIENT PRESENTED ON (B)(6)2024 FOR A PRIMARY PROCEDURE ON TOOTH #15. THE PATIENT RETURNED BEFORE THE SECOND STAGE SURGERY ON (B)(6)2025. AN INFECTION AND INFLAMMATION WERE OBSERVED. THE PROVIDER DETERMINED THAT THE IMPLANT HAD FAILED TO OSSEOINTEGRATE. THE IMPLANT WAS REMOVED AND REPLACED. THE PATIENTS' SYMPTOMS WERE RESOLVED AFTER REMOVAL. NO PERMANENT INJURY OR HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2550076 GLIDEWELL HT IMPLANT Ø5.0 X 10 MM GLIDEWELL HT IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1189-IMP0015 6240877

Patients

Seq Age Sex Outcome Treatment
1 60 YR Male Required Intervention