FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 23305868 · Received October 15, 2025

Report

Report Number
9617229-2025-17656
Event Type
Injury
Date Received
October 15, 2025
Date of Event
September 25, 2025
Report Date
October 15, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT RUPTURE WAS RECEIVED ON OCTOBER 29, 2025 WITH LOT NUMBER 3031011. BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: ¿ RUPTURE: OBSERVED BROKEN DEVICE ASSESSED AS SURGICAL DAMAGE. AS PER THE INVESTIGATION PROCEDURE, CREASES WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "RUPTURE". THE DEVICE WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "RUPTURE". THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2736954 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3031011

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female Required Intervention