FDA Adverse Event Malfunction Summary report: N

NUTRILINE

MDR report key: 23303144 · Received October 15, 2025

Report

Report Number
2245270-2025-00124
Event Type
Malfunction
Date Received
October 15, 2025
Date of Event
May 15, 2025
Report Date
December 15, 2025
Manufacturer
VYGON USA
Product Code
LJS
PMA / PMN Number
K051690
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WE RECEIVED THE CATHETER WITH NEEDLE-FREE DEVICE (NON-VYGON GERMANY GMBH) AS FAULTY SAMPLE. THE CATHETER TUBE WAS SHORTENED AT THE 10 CM MARKING.THE ATTEMPT TO FLUSH THE CATHETER TUBE WITH WATER WAS SUCCESSFUL AFTER CLAMPING THE CATHETER TIP TO INCREASE THE PRESSURE AND A LEAKAGE WAS DETECTED AT THE WING. MICROSCOPIC EXAMINATION OF THE CATHETER TUBE REVEALED A TEAR DIRECTLY AT THE WING. THE FRACTURE SURFACE WAS ROUGH, INDICATING EXCESSIVE TENSILE FORCE AND THE USE OF ALCOHOL-BASED DISINFECTION. IN GENERAL, THERE ARE VARIOUS EVENTS THAT CAN LEAD TO A LEAKING CATHETER: 1. TENSILE FORCE WHICH MAY BE CAUSED BY: DRESSING CHANGE - IN SOME INSTANCES THE CATHETER CAN BECOME ADHERED TO THE DRESSING AND ADDITIONAL PULLING IS REQUIRED TO FREE IT; PLACING STRESS ON THE LINE COULD RESULT IN A TENSILE FRACTURE. FOR BABIES, ROUTINE CARE (WHEN LIFTING THE BABY TO CHANGE THE BEDDING) AND MOVEMENT OF THE BABY ITSELF COULD RESULT IN A TENSILE FRACTURE. 2. MECHANICAL DAMAGE BY A SHARP INSTRUMENT (FOR EXAMPLE SCISSOR, FORCEPS OR SCALPEL) DURING DRESSING CHANGE. 3. ALCOHOL-BASED DISINFECTANT - CONCERNING THIS, THERE ARE SOME WARNINGS IN THE IFU: "BE AWARE THAT ORGANIC SOLVENTS SUCH AS ALCOHOL OR ACETONE MAY INTERACT WITH CATHETER MATERIAL AND WEAKEN IT." IN ADDITION, A MANUFACTURING FAULT CAN BE EXCLUDED AS EACH CATHETER IS FLOW AND LEAK TESTED DURING PRODUCTION, AND THE CATHETER DID NOT LEAK FOR 11 DAYS AS STATED BY THE CUSTOMER ("DWELL TIME: 11 DAYS"). HAVING CHECKED THE BATCH HISTORY RECORDS; NO DEVIATIONS WERE FOUND. THE BATCHES COMPLIED WITH THEIR SPECIFICATIONS AND WERE RELEASED. EACH CATHETER IS FLOW AND LEAK TESTED DURING PRODUCTION. THE TENSILE FORCE AND DIMENSIONS OF CATHETER AND SET COMPONENTS ARE RANDOMLY CHECKED. INCOMING GOODS INSPECTIONS AND TWO 100% VISUAL TESTS AFTER PACKAGING ARE CARRIED OUT. THERE ARE THREE FURTHER COMPLAINTS FOR BATCH 8236275, TWO FURTHER COMPLAINTS FOR BATCH 8181455 AND FIVE FURTHER COMPLAINTS REGARDING A LEAKING CATHETER AT THE WING ON PRODUCT CODE 4G07125231 WITHIN THE LAST THREE YEARS. HOWEVER, NONE OF THESE FURTHER COMPLAINTS IS RELATED TO A MANUFACTURING FAULT. THIS QUERY RELATES TO ALL COMPLAINTS THAT HAVE COME TO OUR ATTENTION WORLDWIDE. NO FURTHER CORRECTIVE ACTION WAS INITIATED BY QUALITY MANAGEMENT AS THERE IS NO HINT OF A MANUFACTURING FAULT.

Additional Manufacturer Narrative · 0

THE MALFUNCTIONING DEVICE WILL BE RETURNED FOR EVALUATION AS PART OF THE COMPLAINT INVESTIGATION. UPON RECEIPT, IT WILL BE INVESTIGATED. THE RESULTS OF THIS INVESTIGATION ARE STILL PENDING AND WILL BE REPORTED TO THE FDA WITHIN THIRTY DAYS OF ITS CONCLUSION VIA A FOLLOW-UP MDR.

Description of Event or Problem · 0

PICC LEAKING NEAR WINGS, PICC WAS REMOVED AND A REPLACEMENT PICC WAS PLACED.

Description of Event or Problem · 0

PICC LEAKING NEAR WINGS, PICC WAS REMOVED AND A REPLACEMENT PICC WAS PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2004749 NUTRILINE INTRAVASCULAR CATHETER, LONG-TERM LJS VYGON USA 1252.31G 24C021D

Patients

Seq Age Sex Outcome Treatment
1 1 MO Male