FDA Adverse Event Malfunction Summary report: N

SONATA

MDR report key: 23297417 · Received October 15, 2025

Report

Report Number
9710014-2025-00952
Event Type
Malfunction
Date Received
October 15, 2025
Date of Event
August 10, 2025
Report Date
November 21, 2025
Manufacturer
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
Product Code
MCM
UDI-DI
09008737049119
PMA / PMN Number
P000025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: BASED ON THE RECEIVED INFORMATION, A DAMAGE TO THE ACTIVE ELECTRODE DUE TO EXCESSIVE MECHANICAL STRESS SEEMS VERY LIKELY. HOWEVER TO DETERMINE AN EXACT ROOT CAUSE DEVICE INVESTIGATION WOULD BE NECESSARY. RE-IMPLANTATION IS CONSIDERED BUT NO DATE HAS BEEN SCHEDULED YET.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: BASED ON THE RECEIVED INFORMATION, A DAMAGE TO THE ACTIVE ELECTRODE DUE TO EXCESSIVE MECHANICAL STRESS SEEMS VERY LIKELY. HOWEVER TO DETERMINE AN EXACT ROOT CAUSE DEVICE INVESTIGATION WOULD BE NECESSARY. RE-IMPLANTATION IS CONSIDERED BUT NO DATE HAS BEEN SCHEDULED YET. CORRECTION: G07003; G02015. B14 WAS DELETED. C070602; C070604.

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN EXPLANTED. IF IT SHOULD BE EXPLANTED, IT SHOULD BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WHEN AVAILABLE, A DEVICE FAILURE ANALYSIS WILL BE SUBMITTED AS A FOLLOW UP REPORT.

Description of Event or Problem · 0

THE USER DOES NOT RESPOND TO SOUND WITH THE DEVICE FOR THE LAST TWO MONTHS. THE USER WILL BE RE-IMPLANTED.

Description of Event or Problem · 0

THE USER DOES NOT RESPOND TO SOUND WITH THE DEVICE FOR THE LAST TWO MONTHS. THE USER WILL BE RE-IMPLANTED.

Description of Event or Problem · 0

THE USER DOES NOT RESPOND TO SOUND WITH THE DEVICE FOR THE LAST 2 MONTHS. THE USER WILL BE RE-IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2004313 SONATA COCHLEAR IMPLANT MCM MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA 09008737049119

Patients

Seq Age Sex Outcome Treatment
1 6 YR Male